The influence of lack of reference conditions on dosimetry in pre-clinical radiotherapy with medium energy x-ray beams.

Anna Subiel1, Ileana Silvestre Patallo, Hugo Palmans, Miriam Barry, Amanda Tulk, Georgios Soultanidis, John Greenman, Victoria L Green, Christopher Cawthorne, Giuseppe Schettino.   

Abstract

Despite well-established dosimetry in clinical radiotherapy, dose measurements in pre-clinical and radiobiology studies are frequently inadequate, thus undermining the reliability and reproducibility of published findings. The lack of suitable dosimetry protocols, coupled with the increasing complexity of pre-clinical irradiation platforms, undermines confidence in preclinical studies and represents a serious obstacle in the translation to clinical practice. To accurately measure output of a pre-clinical radiotherapy unit, appropriate Codes of Practice (CoP) for medium energy x-rays needs to be employed. However, determination of absorbed dose to water (Dw) relies on application of backscatter factor (Bw) employing in-air method or carrying out in-phantom measurement at the reference depth of 2 cm in a full backscatter (i.e. 30 × 30 × 30 cm3) condition. Both of these methods require thickness of at least 30 cm of underlying material, which are never fulfilled in typical pre-clinical irradiations. This work is focused on evaluation the effects of the lack of recommended reference conditions in dosimetry measurements for pre-clinical settings and is aimed at extending the recommendations of the current CoP to practical experimental conditions and highlighting the potential impact of the lack of correct backscatter considerations on radiobiological studies.

Entities:  

Year:  2020        PMID: 32109893     DOI: 10.1088/1361-6560/ab7b30

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Use of calculations to validate beam quality and relative dose measurements for a kilovoltage X-ray therapy unit.

Authors:  B J Healy; R F Hill
Journal:  Phys Eng Sci Med       Date:  2022-04-05

2.  A preclinical radiotherapy dosimetry audit using a realistic 3D printed murine phantom.

Authors:  Emma R Biglin; Adam H Aitkenhead; Gareth J Price; Amy L Chadwick; Elham Santina; Kaye J Williams; Karen J Kirkby
Journal:  Sci Rep       Date:  2022-04-26       Impact factor: 4.996

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.