In vitro comparison of the accuracy of four intraoral scanners and three conventional impression methods for two neighboring implants.

Elena Roig1, Luis Carlos Garza1, Natalia Álvarez-Maldonado1, Paulo Maia2, Santiago Costa1, Miguel Roig1, José Espona1.   

Abstract

PURPOSE: To determine whether the accuracy of two-implant model impressions taken with optical scanners was inferior to that of those taken with elastomeric materials.
MATERIALS AND METHODS: Impressions of a resin reference model with two almost parallel implants were taken using three elastomeric impressions (closed tray technique, open tray nonsplinted technique and open tray splinted technique) and scanned with four optical scanners (CEREC Omnicam, 3M True Definition Scanner, 3Shape TRIOS3 and Carestream CS 3600). STL files of the different methods were superimposed and analyzed with control software (Geomagic Control X, 3D systems) to determine the mean deviation between scans.
RESULTS: Compared to elastomeric impressions, optical impressions showed a significantly improved mean precision. TRIOS3 and CS3600 showed a significantly improved mean trueness compared to that of closed tray, CEREC Omnicam and TrueDefinition. All methods showed a certain degree of implant rotation over their axes, which was significantly higher in the closed tray and the open tray nonsplinted techniques.
CONCLUSIONS: Optical impressions, taken under these in vitro conditions, showed improved accuracy compared with that of elastomeric impressions.

Entities:  

Year:  2020        PMID: 32106275     DOI: 10.1371/journal.pone.0228266

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


  5 in total

1.  Accuracy of commercial intraoral scanners.

Authors:  Mattia Sacher; Georg Schulz; Hans Deyhle; Kurt Jäger; Bert Müller
Journal:  J Med Imaging (Bellingham)       Date:  2021-05-24

2.  Accuracy of full-arch digitalization for partially edentulous jaws - a laboratory study on basis of coordinate-based data analysis.

Authors:  Panagiotis Kontis; Jan-Frederik Güth; Christine Keul
Journal:  Clin Oral Investig       Date:  2022-01-04       Impact factor: 3.573

3.  In vitro comparative analysis of scanning accuracy of intraoral and laboratory scanners in measuring the distance between multiple implants.

Authors:  Reiji Natsubori; Shota Fukazawa; Toyokazu Chiba; Norimasa Tanabe; Hidemichi Kihara; Hisatomo Kondo
Journal:  Int J Implant Dent       Date:  2022-04-13

Review 4.  Trueness and precision of digital implant impressions by intraoral scanners: a literature review.

Authors:  Minoru Sanda; Keita Miyoshi; Kazuyoshi Baba
Journal:  Int J Implant Dent       Date:  2021-07-27

5.  Trueness of 12 intraoral scanners in the full-arch implant impression: a comparative in vitro study.

Authors:  Francesco Guido Mangano; Oleg Admakin; Matteo Bonacina; Henriette Lerner; Vygandas Rutkunas; Carlo Mangano
Journal:  BMC Oral Health       Date:  2020-09-22       Impact factor: 2.757

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.