Diagnostic Performance of Dual-Layer Computed Tomography for Deep Vein Thrombosis in Indirect Computed Tomography Venography.

Shota Tanoue1, Takeshi Nakaura1, Yuji Iyama2, Ayumi Iyama1, Yasunori Nagayama1, Morikatsu Yoshida3, Yasuyuki Yamashita1.   

Abstract

BACKGROUND: The aim of this study was to evaluate the quality and diagnostic performance of virtual monochromatic images (VMI) obtained with dual-layer dual-energy computed tomography (DL-DECT) during indirect CT venography (CTV) for deep vein thrombosis (DVT).Methods and 
Results: This retrospective study was approved by the Institutional Review Board, which waived the requirement for informed consent. We retrospectively enrolled 45 patients who underwent CTV with DL-DECT, and VMI were retrospectively generated. We compared the venous attenuation, noise, contrast, and contrast-to-noise ratio (CNR) between VMI with the highest CNR and conventional CT on paired t-test. Furthermore, we compared the pooled area under the curve (AUC) of each technique with Delong's test in 34 patients who underwent color Doppler ultrasonography. The 40-keV VMI had the best CNR. The noise was significantly lower on 40-keV (9.7±2.5 HU) than on 120-kVp VMI (10.5±2.5 HU; P<0.01). The contrast (120 kVp, 38.2±15.3 HU vs. 40 keV, 131.6±43.6 HU) and CNR (120 kVp, 3.8±1.7 vs. 40 keV, 14.4±6.1) were significantly higher in 40-keV VMI than in 120-kVp VMI (P<0.01). Furthermore, the pooled AUC was significantly higher for 40-keV (0.84) than for 120-kVp VMI (0.78; P=0.03).
CONCLUSIONS: In indirect CTV, 40-keV VMI obtained with DL-DECT offers better image quality and diagnostic performance for DVT than conventional CT.

Entities:  

Keywords:  Phlebography; Spiral computed tomography; Venous thrombosis; Virtual monochromatic image; X-ray energy

Year:  2020        PMID: 32101814     DOI: 10.1253/circj.CJ-19-0722

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  1 in total

1.  Spiral Computed Tomography in the Quantitative Measurement of the Adjacent Structure of the Left Atrial Appendage in Patients with Atrial Fibrillation.

Authors:  Zhen Zhang; Wei Yan
Journal:  J Healthc Eng       Date:  2021-11-30       Impact factor: 2.682

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.