Literature DB >> 32073725

Utility of the long DrySeal sheath in facilitating transcatheter pulmonary valve implantation with the Edwards Sapien 3 valve.

Terunobu Fukuda1, Weiyi Tan1, Soraya Sadeghi1, Jeannette Lin1, Morris Salem2, Daniel Levi1,3, Jamil Aboulhosn1,3.   

Abstract

OBJECTIVE: To assess the utility of the 65-cm-long Gore DrySeal sheath when compared to the standard 36-cm-long Edwards expandable sheath (e-sheath) for transcatheter pulmonary valve implantation (TPVI) with the Edwards Sapien 3 valve.
METHODS: All patients who underwent TPVI with the Sapien 3 valve, excluding those performed via hybrid approach, at our center between September 2015 and November 2019 were retrospectively reviewed and compared between two groups.
RESULTS: A total of 94 patients were enrolled; 29 patients underwent TPVI with the Sapien valve using the DrySeal sheath and 65 underwent TPVI using the e-sheath. The height and body weight of patients implanted using the DrySeal sheath ranged from 137 to 193 cm and from 33 to 129 kg, respectively. Valve delivery time was significantly shorter in the DrySeal group (median time 4 min 33 s vs. 9 min 6 s, p = .002). There were no complications in the DrySeal group (0/27). Nine procedural complications occurred in the e-sheath group (9/65), five of which were potentially directly related to sheath choice, including tricuspid valve injury in four and embolization of the tip of the e-sheath during retrieval of a ruptured balloon in one patient.
CONCLUSIONS: TPVI with the Sapien 3 valve using the 65-cm-long DrySeal sheath facilitates faster and safer valve implantation when compared to the e-sheath.
© 2020 Wiley Periodicals, Inc.

Entities:  

Keywords:  DrySeal sheath; Sapien valve; congenital heart disease; transcatheter pulmonary valve implantation

Year:  2020        PMID: 32073725     DOI: 10.1002/ccd.28776

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  A Modified Buddy-Wire Technique for Crossing of the Interatrial Septum With the Sapien 3 Valve During Transseptal Mitral Valve-in-Valve/Ring Procedures.

Authors:  Sagar N Doshi; Panagiotis Savvoulidis; Anthony Mechery; Ewa Lawton; Peter F Ludman; Adnan Nadir
Journal:  CJC Open       Date:  2022-07-15
  1 in total