Transcutaneous bilirubinometry. Evaluation of accuracy and reliability in a large population.

Y Yamauchi1, I Yamanouchi.   

Abstract

A total of 576 transcutaneous bilirubin measurements were performed on 336 Japanese full-term breast-fed newborn infants during the first twelve days of life. Our present study revealed that transcutaneous bilirubin measurements obtained from the forehead, chest, and sternum correlated well with serum bilirubin concentrations measured by AO bilirubinometer (0.910-0.922, p less than 0.001, n = 576). The 95% confidence limits were +/- 3.04 mg/dl for the forehead, +/- 2.85 mg/dl for the chest, and +/- 2.84 mg/dl for the sternum readings. The overall mean of values from the forehead, chest and sternum, when compared with individual means, was found to correlate better with serum bilirubin concentrations (r = 0.930, p less than 0.001, n = 576) and improve the 95% confidence limits to +/- 2.68 mg/dl. These results demonstrated that the accuracy and reliability of TcB measurement could be increased further with multiple site measurement. The study clearly indicates that transcutaneous bilirubinometry is useful for clinical screening of serum bilirubin levels in Japanese full-term newborn infants.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3207014     DOI: 10.1111/j.1651-2227.1988.tb10757.x

Source DB:  PubMed          Journal:  Acta Paediatr Scand        ISSN: 0001-656X


  5 in total

1.  Assessment of neonatal jaundice in low birth weight infants comparing transcutaneous, capillary and arterial bilirubin levels.

Authors:  M Amato; P Huppi; D Markus
Journal:  Eur J Pediatr       Date:  1990-11       Impact factor: 3.183

2.  Transcutaneous bilirubin in predicting hyperbilirubinemia in term neonates.

Authors:  Y Ramesh Bhat; Amitha Rao
Journal:  Indian J Pediatr       Date:  2008-02       Impact factor: 1.967

3.  Transcutaneous bilirubinometry in neonatal intensive care units.

Authors:  A Knudsen; F Ebbesen
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1996-07       Impact factor: 5.747

4.  A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields.

Authors:  C Mae Wong; P J E van Dijk; I A Laing
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-09       Impact factor: 5.747

5.  Comparison of a new transcutaneous bilirubinometer (Bilimed) with serum bilirubin measurements in preterm and full-term infants.

Authors:  Tanja Karen; Hans Ulrich Bucher; Jean-Claude Fauchère
Journal:  BMC Pediatr       Date:  2009-11-12       Impact factor: 2.125

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.