| Literature DB >> 32067431 |
In Young Yoo1, Ji Youn Kim2, Young Kyung Yoon3,4, Hee Jae Huh5, Nam Yong Lee1.
Abstract
The recent increase in severe fever with thrombocytopenia syndrome (SFTS) cases has led to the development of the SFTS-QS kit (MiCoBioMed, Seongnam, Korea) for detecting the SFTS virus (SFTSV, now renamed Huaiyangshan banyangvirus). SFTS-QS is a qualitative real-time reverse transcription PCR assay based on lab-on-a-chip technology. We evaluated the performance of the SFTS-QS kit and compared it with that of the PowerChek SFTSV Real-time PCR kit (PowerChek; Kogene Biotech, Seoul, Korea). A total of 117 serum samples were simultaneously assayed using the SFTS-QS and PowerChek kits. Sanger sequencing targeting the S and M segments of SFTSV was performed as the reference method. The total turnaround time of the two kits was compared. The SFTS-QS results agreed with those of PowerChek with a kappa value of 0.92. The diagnostic sensitivity and specificity of the SFTS-QS kit were both 100% (14/14 and 103/103, respectively), whereas those of the PowerChek kit were 100% (14/14) and 98.1% (101/103), respectively. The results of SFTS-QS and PowerChek were comparable; however, the SFTS-QS kit required a shorter total turnaround time. The SFTS-QS kit produced accurate and fast results and thus could serve as a useful tool for detecting SFTSV. © The Korean Society for Laboratory Medicine.Entities:
Keywords: Huaiyangshan banyangvirus; Performance; PowerChek SFTSV Real-time PCR kit; SFTS virus; SFTS-QS kit; Turnaround time
Mesh:
Substances:
Year: 2020 PMID: 32067431 PMCID: PMC7054697 DOI: 10.3343/alm.2020.40.4.317
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Performance of the real-time RT-PCR kits for SFTSV detection
| Method | Reference method (RT-PCR and sequencing) | Accuracy % (95% CI) | ||
|---|---|---|---|---|
| Positive | Negative | |||
| SFTS-QS Kit | Positive | 14 | 0 | Sensitivity 100% (73.2%–100%) |
| Negative | 0 | 103 | Specificity 100% (95.5%–100%) | |
| PowerChek SFTSV Real-time PCR Kit | Positive | 14 | 2 | Sensitivity 100% (73.2%–100%) |
| Negative | 0 | 101 | Specificity 98.1% (92.5%–99.7%) | |
| Agreement % (95% CI) | PPA 87.5% (95% CI, 60.4%–97.8%) | |||
| NPA 100% (95% CI, 95.4%–100%) | ||||
| Kappa value 0.92 (95% CI, 0.84–1.00) | ||||
Abbreviations: RT-PCR, reverse transcription PCR; SFTSV, severe fever with thrombocytopenia syndrome virus; CI, confidence interval, PPA, positive percent agreement, NPV negative percent agreement.