Fahad Alkhudhairy1, Sulieman S Al-Johany2, Mustafa Naseem3, Mohammed Bin-Shuwaish4, Fahim Vohra5.
Abstract
OBJECTIVE: The purpose of the present study was to assess dentin shear bond strength (SBS) and mode of bond failure of bioactive cement (BA) in comparison to conventional resin cement when photosensitized by Er,Cr: YSGG Laser (ECL).Entities:
Keywords: Bioactive cement; Conventional resin cement; Dental aesthetics; Er,Cr:YSGG; Shear bond strength
Year: 2020 PMID: 32063937 PMCID: PMC6994876 DOI: 10.12669/pjms.36.2.1284
Source DB: PubMed Journal: Pak J Med Sci ISSN: 1681-715X Impact factor: 1.088
Types of luting cement along with composition used in the study.
| Product | Type of polymerization | Composition | Manufacturer |
|---|---|---|---|
| Bioactive Cement | Dual Cured | Diurethane and other methacrylates, polyacrylic acid, amorphous silica, sodium and fluoride | Pulpdent |
| Variolink II Aesthetic cement | Dual Cured | UDMA, TEGDMA, Self-curing initiators silicon dioxide, pigment, stabilizers, light curing initiators | Ivocalar, Vivadent |
| Calibra luting cement | Dual Cured | BISGMA monomer, Benzoyl peroxide Camphorquinone, coupling agent, glass filler, peroxide | Dentsply, Caulk |
| Maxcem Elite cement | Dual Cured | Proprietary redox initiators, photo initiators, Resin: GPDMs, DMAs Fillers: barium, fluor- aluminosilicate, and silica | Kerr, Corporation |
Means and SD for bond strength among different study groups using ANOVA and Tukey multiple comparisons test.
| Surface treatment /type of cement used in experimental groups | Mean ± SD (Mpa) | P-value! |
|---|---|---|
| Group-1 ECL-BA | 15.48±3.62 | <0.001 |
| Group-2 ECL-V ≠ | 20.74±4.15 | |
| Group-3 ECL-ME | 14.25±3.55 | |
| Group-4 ECL-C ≠ | 21.55±3.08 |
ECL: Er,Cr:YSGG laser, BA: Bioactive Cement, V: Variolink II, ME: Maxcem Elite, C: Calibra The highest and lowest SBS values are in bold, ≠ Significantly different from groups- ECL-BA, ECL-ME (p <0.05),
Significantly different from groups- ECL-V, ECL-C (p <0.05), (Tukey multiple comparison test), ! Showing significant difference among study group (ANOVA).
Modes of failure among different experimental groups
| Experimental groups | Adhesive (%) | Cohesive (%) | Admixed (%) |
|---|---|---|---|
| Group-1 ECL-BA | 90 | 10 | - |
| Group-2 ECL-V | 20 | 10 | 70 |
| Group-3 ECL-ME | 70 | 20 | 10 |
| Group-4 ECL-C | 30 | 10 | 60 |
ECL: Er,Cr:YSGG laser, BA: Bioactive Cement, V: Variolink II, ME: Maxcem Elite, C: Calibra.