Literature DB >> 32051948

Major haemorrhage following vascular injury during exchange of cardiac pacemaker leads.

A Elrefaey1, A Pai2.   

Abstract

An 80-year-old lady suffered loss of cardiac output during an elective pacemaker lead exchange. Right ventricular perforation, cardiac tamponade, vascular lacerations and circulatory shock are known major complications of pacemaker lead exchange and may necessitate aggressive resuscitation. An emergency sternotomy revealed the administered fluid, blood products and resuscitative drugs, which were administered through an upper limb cannula, had likely extravasated at the site of an iatrogenic injury to the superior vena cava. Unfortunately, further resuscitation attempts were unsuccessful. There are no recommendations regarding sites for intravascular access during pacemaker lead exchange and we argue that lower limb veins should be considered for venous access, as the superior vena cava and innominate veins are possible sites of unintentional vascular injury.
© 2019 Association of Anaesthetists.

Entities:  

Keywords:  cardiac pacemaker; intravascular access; major haemorrhage; vascular injury

Year:  2019        PMID: 32051948      PMCID: PMC6931285          DOI: 10.1002/anr3.12016

Source DB:  PubMed          Journal:  Anaesth Rep        ISSN: 2637-3726


  6 in total

Review 1.  Extraction of cardiac rhythm devices: indications, techniques and outcomes for the removal of pacemaker and defibrillator leads.

Authors:  F M Farooqi; S Talsania; S Hamid; C A Rinaldi
Journal:  Int J Clin Pract       Date:  2010-07       Impact factor: 2.503

2.  Peri-operative extracorporeal cardiopulmonary resuscitation.

Authors:  M Charlesworth; A Ashworth
Journal:  Anaesthesia       Date:  2019-02       Impact factor: 6.955

Review 3.  2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction.

Authors:  Fred M Kusumoto; Mark H Schoenfeld; Bruce L Wilkoff; Charles I Berul; Ulrika M Birgersdotter-Green; Roger Carrillo; Yong-Mei Cha; Jude Clancy; Jean-Claude Deharo; Kenneth A Ellenbogen; Derek Exner; Ayman A Hussein; Charles Kennergren; Andrew Krahn; Richard Lee; Charles J Love; Ruth A Madden; Hector Alfredo Mazzetti; JoEllyn Carol Moore; Jeffrey Parsonnet; Kristen K Patton; Marc A Rozner; Kimberly A Selzman; Morio Shoda; Komandoor Srivathsan; Neil F Strathmore; Charles D Swerdlow; Christine Tompkins; Oussama Wazni
Journal:  Heart Rhythm       Date:  2017-09-15       Impact factor: 6.343

Review 4.  Surgical management of major intrathoracic hemorrhage resulting from high-risk transvenous pacemaker/defibrillator lead extraction.

Authors:  Justine M Caniglia-Miller; Walter D Bussey; Natalie M Kamtz; Shane F Tsai; Christopher C Erickson; Daniel R Anderson; Michael J Moulton
Journal:  J Card Surg       Date:  2014-12-29       Impact factor: 1.620

5.  Extracorporeal cardiopulmonary resuscitation in refractory intra-operative cardiac arrest: an observational study of 12-year outcomes in a single tertiary hospital.

Authors:  J J Min; C K Tay; D K Ryu; W Wi; K Sung; Y T Lee; Y H Cho; J-H Lee
Journal:  Anaesthesia       Date:  2018-09-26       Impact factor: 6.955

6.  Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction.

Authors:  Robert G Hauser; William T Katsiyiannis; Charles C Gornick; Adrian K Almquist; Linda M Kallinen
Journal:  Europace       Date:  2009-11-27       Impact factor: 5.214