Polyether polyurethanes: biostable or not?

K B Stokes1.   

Abstract

Certain polyether polyurethanes have been shown to be biostable in long-term implant studies. Others retain good bulk properties, but have been shown to develop cracks on their tissue contacting surfaces. Two cracking mechanisms have been identified, in vivo stress cracking and metal ion oxidation. Stress cracking is the result of an interaction between the in vivo mammalian environment and residual stress (strain) in the implanted polymer. Mild autooxidation can be initiated by stress cracking. More extensive autooxidation can be initiated and propagated by corrosion of metallic device components, especially the corrosion products of cobalt. Both mechanisms are controllable, thus, do not necessarily preclude the use of polyether polyurethanes in implantable devices.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3204495     DOI: 10.1177/088532828800300205

Source DB:  PubMed          Journal:  J Biomater Appl        ISSN: 0885-3282            Impact factor:   2.646


  4 in total

Review 1.  Tissue-engineered heart valve: future of cardiac surgery.

Authors:  Radoslaw A Rippel; Hossein Ghanbari; Alexander M Seifalian
Journal:  World J Surg       Date:  2012-07       Impact factor: 3.352

2.  Degradation of biomaterials by phagocyte-derived oxidants.

Authors:  K Sutherland; J R Mahoney; A J Coury; J W Eaton
Journal:  J Clin Invest       Date:  1993-11       Impact factor: 14.808

3.  A comparison of the shock-absorbing properties of cervical disc prosthesis bearing materials.

Authors:  Michael C Dahl; Stephen Jacobsen; Newton Metcalf; Rick Sasso; Randal P Ching
Journal:  SAS J       Date:  2011-06-01

4.  In Vivo Stability of Polyurethane-Based Electrospun Vascular Grafts in Terms of Chemistry and Mechanics.

Authors:  Alexander A Gostev; Inna K Shundrina; Vitaliy I Pastukhov; Alexey V Shutov; Vera S Chernonosova; Andrey A Karpenko; Pavel P Laktionov
Journal:  Polymers (Basel)       Date:  2020-04-07       Impact factor: 4.967

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.