Time delay of the qCON monitor and its performance during state transitions.

Robert Zanner1,2, Gerhard Schneider1, Adrian Meyer2, Eberhard Kochs1, Matthias Kreuzer3.   

Abstract

We investigated the performance of the qCON index regarding its time delay for sudden changes in the anesthetic level as well as to separate responsiveness from unresponsiveness during loss and return of responsiveness (LOR and ROR). For evaluation of the time delay, we replayed relevant EEG episodes to the qCON to simulate sudden changes between the states (i) awake/sedation, (ii) adequate anesthesia, or (iii) suppression. We also replayed EEG from 40 patients during LOR and ROR to evaluate the qCON's ability to separate responsiveness from unresponsiveness. The time delays depended on the type of transition. The delays for the important transition between awake/sedation and adequate anesthesia were 21(5) s from awake/sedation to adequate anesthesia and 26(5) s in the other direction. The performance of the qCON to separate responsiveness from unresponsiveness depended on signal quality, the investigation window, i.e. ± 30 s or ± 60 s around LOR/ROR, and the specific transition being tested. AUC was 0.63-0.90 for LOR and 0.61-0.79 for ROR. Time delay and performance during state transitions of the qCON were similar to other monitoring systems such as bispectral index. The better performance of qCON during LOR than ROR probably reflects the sudden change in EEG activity during LOR and the more heterogeneous EEG during ROR.

Entities:  

Keywords:  Anesthesia, general; Electroencephalography; Humans; Monitoring

Year:  2020        PMID: 32040794     DOI: 10.1007/s10877-020-00480-4

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  12 in total

1.  Computation of measures of effect size for neuroscience data sets.

Authors:  Harald Hentschke; Maik C Stüttgen
Journal:  Eur J Neurosci       Date:  2011-11-14       Impact factor: 3.386

Review 2.  A program for computing the prediction probability and the related receiver operating characteristic graph.

Authors:  Denis Jordan; Marcel Steiner; Eberhard F Kochs; Gerhard Schneider
Journal:  Anesth Analg       Date:  2010-11-08       Impact factor: 5.108

3.  Does the cerebral state index separate consciousness from unconsciousness?

Authors:  Stefanie Pilge; Jasmin Blum; Eberhard F Kochs; Stephan-Andreas Schöniger; Matthias Kreuzer; Gerhard Schneider
Journal:  Anesth Analg       Date:  2011-10-14       Impact factor: 5.108

Review 4.  Unresponsiveness ≠ unconsciousness.

Authors:  Robert D Sanders; Giulio Tononi; Steven Laureys; Jamie W Sleigh
Journal:  Anesthesiology       Date:  2012-04       Impact factor: 7.892

5.  Comparison of the qCON and qNOX indices for the assessment of unconsciousness level and noxious stimulation response during surgery.

Authors:  Umberto Melia; Eva Gabarron; Mercé Agustí; Nuria Souto; Patricia Pineda; Joan Fontanet; Montserrat Vallverdu; Erik Weber Jensen; Pedro Gambus
Journal:  J Clin Monit Comput       Date:  2016-10-20       Impact factor: 2.502

6.  Monitoring depth of sedation: evaluating the agreement between the Bispectral Index, qCON and the Entropy Module's State Entropy during flexible bronchoscopy.

Authors:  Johannes N Müller; Matthias Kreuzer; Paul S García; Gerhard Schneider; Hubert Hautmann
Journal:  Minerva Anestesiol       Date:  2017-02-08       Impact factor: 3.051

Review 7.  General anesthesia, sleep, and coma.

Authors:  Emery N Brown; Ralph Lydic; Nicholas D Schiff
Journal:  N Engl J Med       Date:  2010-12-30       Impact factor: 91.245

8.  Monitoring hypnotic effect and nociception with two EEG-derived indices, qCON and qNOX, during general anaesthesia.

Authors:  E W Jensen; J F Valencia; A López; T Anglada; M Agustí; Y Ramos; R Serra; M Jospin; P Pineda; P Gambus
Journal:  Acta Anaesthesiol Scand       Date:  2014-07-04       Impact factor: 2.105

9.  Anesthesia awareness and the bispectral index.

Authors:  Michael S Avidan; Lini Zhang; Beth A Burnside; Kevin J Finkel; Adam C Searleman; Jacqueline A Selvidge; Leif Saager; Michelle S Turner; Srikar Rao; Michael Bottros; Charles Hantler; Eric Jacobsohn; Alex S Evers
Journal:  N Engl J Med       Date:  2008-03-13       Impact factor: 91.245

10.  New findings and trends for depth of anesthesia monitoring.

Authors:  Erik Weber Jensen
Journal:  Korean J Anesthesiol       Date:  2018-10-01
View more
  1 in total

1.  Comparison of Quantium Consciousness Index and Richmond Agitation Sedation Scale in Mechanically Ventilated Critically Ill Patients: An Observational Study.

Authors:  Makam S Harsha; Pradeep K Bhatia; Ankur Sharma; Priyanka Sethi
Journal:  Indian J Crit Care Med       Date:  2022
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.