| Literature DB >> 32036498 |
Carla Maria Irene Quarato1, Mariapia Venuti2, Marco Sperandeo3.
Abstract
Many works in the literature have shown that the increase in the number of B lines is a nonspecific sign of underlying pulmonary disease. Actually these artifacts are the result of a physical effect of ultrasound between the chest wall and the pulmonary air. Nevertheless the intra- and inter-operator variability in B-lines counting does not only reside only in the count itself but depends also on the type and frequency of the probe used, as well as the ultrasound scan machine setting and the patient's chest shape. In our opinion, proposing a software algorithm to count lines B seems like an unproductive effort.Entities:
Keywords: B-lines; Inter-observer reliability; Intra-observer reliability; Transthoracic ultrasound; Ultrasound artifacts
Mesh:
Year: 2020 PMID: 32036498 DOI: 10.1007/s10877-020-00484-0
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502