Flagging performance of Sysmex XN-10 haematology analyser for malaria detection.

Cécile Dumas1, Pauline Tirard-Collet2, Fanélie Mestrallet3, Sandrine Girard3, Laurent Jallades4, Stéphane Picot2,5, Anne-Lise Bienvenu6,5.   

Abstract

AIM: The aim was to assess the flagging performance of Sysmex XN-10 haematology analyser for malaria detection through the parasitic red blood cell ('pRBC') alarm.
METHODS: We retrospectively studied 584 blood samples performed on the Sysmex XN-10 analyser that were tested for malaria. Sensitivity, specificity, positive and negative predictive values, and prevalence were established for the pRBC alarm.
RESULTS: Sensitivity, specificity, and positive and negative predictive values for the pRBC flag were 7.8%, 100%, 100% and 87.7%, respectively. The prevalence of pRBC flag of 0.026% in the overall population was significantly different from the prevalence of 1.027% in the population tested for malaria.
CONCLUSIONS: Considering the excellent specificity and the low prevalence of the flag in the overall population, we suggest, in case of the presence of pRBC flag, the implementation of a rapid review of the blood smear looking for Plasmodium, mostly if the patient had fever and had not been tested for malaria. © Author(s) (or their employer(s)) 2020. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  automated screening; automation; haematology; malaria

Year:  2020        PMID: 32034055     DOI: 10.1136/jclinpath-2019-206382

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  2 in total

1.  Diagnostic accuracy of fluorescence flow-cytometry technology using Sysmex XN-31 for imported malaria in a non-endemic setting.

Authors:  Stéphane Picot; Thomas Perpoint; Christian Chidiac; Alain Sigal; Etienne Javouhey; Yves Gillet; Laurent Jacquin; Marion Douplat; Karim Tazarourte; Laurent Argaud; Martine Wallon; Charline Miossec; Guillaume Bonnot; Anne-Lise Bienvenu
Journal:  Parasite       Date:  2022-05-31       Impact factor: 3.020

2.  Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.

Authors:  Shubham Rastogi; Sukesh Chandran Nair; Pandiyan Murugan; Asady Sukanya Sukumar; Joy J Mammen; Saravanan Mullai
Journal:  Pract Lab Med       Date:  2022-04-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.