Literature DB >> 32029390

How Should I Manage a Challenging Aortic Prosthesis Angle During Valve-in-Valve Implantation?

Michael Bianco1, Abdul Rahman Omer1, Felice Reitknecht1, Edo Kaluski2.   

Abstract

Valve-in-valve TAVR procedure stalled due to challenging aortic to prosthesis angle not allowing the advancement of the Evolut R into the failing 25 mm Magna prosthesis. Efforts to better align the TAVR device with the valve prosthesis by using conventional methods failed. Eventually, a buddy wire allowed advancement and successful deployment of the Evolut R 26 mm within the stenosed surgical valve. The role and methodology of buddy wire use in TAVR, valve-in-valve procedures (first report) and structural heart procedures is reviewed.
Copyright © 2020 Elsevier Inc. All rights reserved.

Keywords:  Bioprosthetic aortic valve stenosis; Buddy wire; TAVR; Valve-in-valve implantation

Year:  2020        PMID: 32029390     DOI: 10.1016/j.carrev.2020.01.018

Source DB:  PubMed          Journal:  Cardiovasc Revasc Med        ISSN: 1878-0938


  1 in total

1.  Description of the Step-by-Step Technique With Snare Catheter for TAVR in Horizontal Aorta.

Authors:  Manuel A Espinoza Rueda; Roberto Muratalla González; Juan F García García; Julieta D Morales Portano; Marco A Alcántara Meléndez; Arnoldo S Jiménez Valverde; Ronald E Rivas Gálvez; Jorge L Campos Delgadillo; César L González; José R Gayosso Ortiz; José A Merino Rajme
Journal:  JACC Case Rep       Date:  2021-12-01
  1 in total