| Literature DB >> 32029390 |
Michael Bianco1, Abdul Rahman Omer1, Felice Reitknecht1, Edo Kaluski2.
Abstract
Valve-in-valve TAVR procedure stalled due to challenging aortic to prosthesis angle not allowing the advancement of the Evolut R into the failing 25 mm Magna prosthesis. Efforts to better align the TAVR device with the valve prosthesis by using conventional methods failed. Eventually, a buddy wire allowed advancement and successful deployment of the Evolut R 26 mm within the stenosed surgical valve. The role and methodology of buddy wire use in TAVR, valve-in-valve procedures (first report) and structural heart procedures is reviewed.Keywords: Bioprosthetic aortic valve stenosis; Buddy wire; TAVR; Valve-in-valve implantation
Year: 2020 PMID: 32029390 DOI: 10.1016/j.carrev.2020.01.018