| Literature DB >> 32014036 |
Maud Salmona1,2, Karl Stefic3,4, Nadia Mahjoub3, Flore Sicre de Fontbrune5, Sarah Maylin3, François Simon6,3, Catherine Scieux6,3, Gérard Socié6,5, Marie-Christine Mazeron6,3, Jérôme LeGoff6,3.
Abstract
BACKGROUND: Standardized and sensitive assays for Epstein Barr Virus (EBV) are needed to define universal cutoff for treatment initiation in allogeneic hematopoietic stem cells transplant recipients. In a context of accreditation and the availability of EBV international standard, we evaluated the Abbott RealTime EBV (RT) assay for EBV quantification in whole blood.Entities:
Keywords: EBV; Hematopoietic stem cell transplant; Monitoring; Quantitative PCR; Rituximab
Year: 2020 PMID: 32014036 PMCID: PMC6998838 DOI: 10.1186/s12985-020-1285-7
Source DB: PubMed Journal: Virol J ISSN: 1743-422X Impact factor: 4.099
Lower limit of detection (LOD) of the Abbott RealTime EBV assay for whole blood
| Expected value (IU/mL) | No. of replicates | Mean value (IU/mL) | Detection rate (%) |
|---|---|---|---|
| 500 | 10 | 1239 | 100 |
| 100 | 10 | 328 | 100 |
| 20 | 10 | 73 | 60 |
The LOD was estimated by using serial dilutions of WHO international standard in whole blood
Quantification of EBV DNA with Abbott RealTime EBV assay in QCMD 2016 EBV samples
| Sample | QCMD results | RT EBV assay results | |||||
|---|---|---|---|---|---|---|---|
| EBV load | n | Range | EBV load | Delta log10 | |||
| 2016 panel 1 | C | 16C1–01 | 3.673 | 52/52 | 2.937–4.772 | 3.819 | 0.146 |
| C | 16C1–02 | 3.200 | 48/52 | 1.398–4.094 | 2.757 | −0.443 | |
| C | 16C1–03 | ND | N/A | N/A | ND | ND | |
| E | 16C1–04 | 2.389 | 37/52 | 1.000–3.732 | ND | ND | |
| C | 16C1–05 | 3.673 | 52/52 | 3.247–4.580 | 3.927 | 0.254 | |
| 2016 panel 2 | C | 16C2–01 | 4.144 | 53/53 | 3.511–4.710 | 4.096 | −0.048 |
| C | 16C2–02 | 2.949 | 49/53 | 2.068–3.781 | 3.074 | 0.125 | |
| C | 16C2–03 | ND | N/A | N/A | ND | ND | |
| E | 16C2–04 | 2.430 | 31/53 | 1.079–3.534 | ND | ND | |
| C | 16C2–05 | 3.684 | 53/53 | 3.117–4.302 | 3.539 | −0.145 | |
| 2017 panel 1 | C | 17C1–01 | 4.112 | 46/47 | 3.002–5.105 | 3.941 | −0.171 |
| E | 17C1–02 | 2.457 | 30/47 | 1.613–3.182 | Pos NQ | ND | |
| C | 17C1–03 | 4.103 | 46/47 | 3.548–5.240 | 3.994 | −0.109 | |
| C | 17C1–04 | 3.573 | 44/47 | 2.444–4.454 | 3.435 | −0.138 | |
| E | 17C1–05 | 2.597 | 27/47 | 1.519–3.872 | ND | ND | |
| 2017 panel 2 | C | 17C2–01 | 4.220 | 50/50 | 3.609–4.837 | 4.179 | −0.041 |
| C | 17C2–02 | 3.661 | 50/50 | 3.029–4.754 | 3.497 | −0.164 | |
| – | 17C2–03 | Excluded | Excluded | ND | |||
| C | 17C2–04 | 3.660 | 49/50 | 3.208–4.089 | 3.609 | −0.051 | |
| C | 17C2–05 | 3.009 | 47/50 | 2.107–3.877 | 3.079 | 0.07 | |
Consensus log10 virus concentrations were calculated as the mean values reported from positives datasets (n) submitted by clinical laboratories
ND, EBV DNA was not detected
Pos NQ, EBV DNA was detected but not quantified
C core, E educational
Excluded, panel member was excluded of the QCMD report
delta log10, log10 copies/ml difference between the RT assay and QCMD consensus
Comparison Of V1 And Rt Ebv Assays On 282 Whole Blood Samples
| RT EBV assay | |||||
|---|---|---|---|---|---|
| Quantified | Detected | Not Detected | Total | ||
| V1 assay | Quantified | 161 | 8 | 4 | 173 |
| Detected | 12 | 5 | 9 | 26 | |
| Not Detected | 3 | 7 | 73 | 83 | |
| Total | 176 | 20 | 86 | 282 | |
Quantified: EBV DNA detected and above the limit of quantification of the assay
Detected: EBV DNA detected and below the limit of quantification of the assay
Fig. 1Correlation plot of EBV DNA load values measured by the two assays. Correlation plot of EBV DNA load values measured by EBV PCR kit v1 and the Abbott RealTime EBV assays for samples quantified with the two assays (n = 161). EBV DNA loads are expressed in log10 IU/mL. For V1 assay, log10 IU/mL were calculated from log copies/mL using previously calculated conversion factor
Fig. 2Bland-Altman plot of the two assays. Bland-Altman plot of the EBV PCR kit v1 and the Abbott RealTime EBV assays for samples quantified with the two assays (n = 161). EBV DNA loads are expressed in log10 IU/mL
Fig. 3Kinetics of EBV DNA before and after rituximab injection measured with the two assays. Kinetics of EBV DNA was measured in 11 hematopoietic stem cell transplant recipients (#1 to #11) with the EBV PCR kit V1 and the Abbott RealTime EBV assay. EBV DNA loads are expressed in log10 IU/mL. Black arrows indicate time of rituximab injection. Detectable but non quantifiable samples are represented as the half of the LLQ of the technique. Non detectable samples are represented as zero on the graph