| Literature DB >> 32010706 |
Laura M Markodimitraki1,2, Inge Stegeman1,2, Adriana L Smit1,2, Hans G X M Thomeer1,2.
Abstract
Objectives: Postoperative follow-up after cochlear implantation lacks a reliable screening method to detect cochlear implant receiver device migration. This study aims to validate a clinically applicable method to assess the position and migration of the cochlear implant receiver device. Study design: Validation study. Setting: Tertiary university medical center. Participants and method: To assess the cochlear implant receiver device location, round markers representing the external magnet were placed on both sides of the head of volunteers. Four independent clinicians took measurements of the distances between reference points on the head and the center of the marker. The reference points were: the lateral canthus (LC), tragus tip (TT), the mastoid angle (MA), and the mandibular angle (AM). Main outcome measures: The inter-clinician reliability was determined by calculating the intraclass correlation coefficient (ICC) and confidence interval (CI) with a two-way mixed model and both consistency and absolute agreement types for each distance.Entities:
Keywords: cochlea; cochlear implantation; cochlear implants; neurotology; pilot projects; sensorineural hearing loss; validation studies
Year: 2020 PMID: 32010706 PMCID: PMC6974530 DOI: 10.3389/fsurg.2019.00078
Source DB: PubMed Journal: Front Surg ISSN: 2296-875X
Figure 1An adhesive marker was placed postero-superiorly of the ear. The numbers represent the reference points: 1, lateral canthus (LC); 2, the tip of the tragus (TT); 3, the mastoid angle (MA); and 4, the mandibular angle (AM). Written consent was obtained from the individual for the use of this image.
Measurements per distance (mm) for all raters and ratios of measurement.
| A | 147.5 | 88.5 | 94.0 | 134.0 | 85.0 |
| B | 148.0 | 88.5 | 93.0 | 140.0 | 84.0 |
| C | 147.5 | 89.0 | 92.5 | 137.5 | 85.0 |
| D | 145.5 | 88.0 | 89.5 | 126.7 | 79.5 |
LC, lateral canthus; TT, tragus tip; MA, mastoid angle; AM, mandibular angle; SD, standard deviation.
Range of measurements per distance (mm) and intra-class correlation coefficient with 95% confidence interval.
| LC to marker | 148.0 (130–169) | 0.90 | [0.80, 0.96] | 0.87 | [0.76, 0.93] |
| TT to marker | 89.0 (75–100) | 0.83 | [0.69, 0.93] | 0.74 | [0.55, 0.88] |
| MA marker | 92.0 (65–111) | 0.75 | [0.56, 0.89] | 0.65 | [0.44, 0.83] |
| AM to marker | 135.0 (113–158) | 0.29 | [0.05, 0.59] | 0.26 | [0.04, 0.55] |
| LC to TT | 84.5 (73–92) | 0.50 | [0.25, 0.74] | 0.47 | [0.24, 0.71] |
LC, lateral canthus; TT, tragus tip; MA, mastoid angle; AM, mandibular angle; SD, standard deviation; ICC, intra-class correlation coefficient; CI, confidence interval.