| Literature DB >> 32003803 |
Yunus C Yalcin1,2, Claudette Kooij1, Dominic A M J Theuns2, Alina A Constantinescu1, Jasper J Brugts1, Olivier C Manintveld1, Sing-Chien Yap1, Tamas Szili-Torok1, Ad J J C Bogers2, Kadir Caliskan1.
Abstract
AIMS: To investigate the prevalence of electromagnetic interference (EMI) between left ventricular assist devices (LVADs) and implantable cardioverter-defibrillators (ICDs)/pacemakers (PMs). METHODS ANDEntities:
Keywords: Electromagnetic interference; End-stage heart failure; Implantable cardioverter-defibrillator; Left ventricular assist device; Pacemaker
Mesh:
Year: 2020 PMID: 32003803 PMCID: PMC7132535 DOI: 10.1093/europace/euaa006
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Electromagnetic interferences between implantable cardioverter-defibrillators and left ventricular assist devices observed in study population
| Patient | ICD manufacturer | Type | LVAD type | Indication | Support duration | Type of EMI | Intervention |
|---|---|---|---|---|---|---|---|
| 1 | St Jude Medical | Unify (CRT-D) | HMII | Bridge-to-transplantation | 25 days | Interference with telemetry function | RF use |
| 2 | St Jude Medical | Elipse (ICD) | HMII | Bridge-to-transplantation | 203 days | Interference with telemetry function | No interrogation possible |
| 3 | St Jude Medical | Atlas (ICD) | HMII | Bridge-to-transplantation | 287 days | Interference with telemetry function | No interrogation possible |
| 4 | St Jude Medical | Fortify (ICD) | HMII | Bridge-to-transplantation | 970 days | Dysfunction ICD system | Device replacement |
| 5 | St Jude Medical | Promote (CRT-D) | HMII | Bridge-to-transplantation | 1061 days | Interference with telemetry function | RF use |
| 6 | St Jude Medical | Promote (CRT-D) | HMII | Bridge-to-transplantation | 1886 days | Interference with telemetry function | RF use |
| 7 | Biotronik | Ilesto (CRT-D) | HM3 | Bridge-to-decision | 199 days | Interference with telemetry function | Patient in lying position |
| 8 | Medtronic | Viva (CRT-D) | HM3 | Bridge-to-transplantation | 275 days | Interference with telemetry function | No interrogation possible |
| 9 | Biotronik | Iperia (CRT-D) | HM3 | Destination therapy | 348 days | Interference with telemetry function | Uplift device |
| 10 | Biotronik | Ilivia (CRT-D) | HM3 | Bridge-to-transplantation | 724 days | Interference with telemetry function | No interrogation possible |
| 11 | Biotronik | Lumax (ICD) | HM3 | Bridge-to-transplantation | 997 days | Interference with telemetry function | Patient in lying position |
CRT-D, cardiac resynchronization therapy defibrillator; EMI, electromagnetic interference; HM, HeartMate; ICD, implantable cardioverter-defibrillator; LVAD, left ventricular assist device; RF, radiofrequency.
After implantation of the LVAD, the ICD device appeared to be in back-up mode (VVI 65 bpm) and shock therapy was disabled. After a multidisciplinary consultation (together with Abbott), it was decided to replace the ICD device. The cause of this problem was attributed to the EMI generated by the implanted HMII. Following the replacement with a device from a different manufacturer, no further issues were observed.