Accuracy assessment of bolus and basal rate delivery of different insulin pump systems used in insulin pump therapy of children and adolescents.

Ralph Ziegler1, Delia Waldenmaier2, Ulrike Kamecke2, Jochen Mende2, Cornelia Haug2, Guido Freckmann2.   

Abstract

BACKGROUND: Continuous subcutaneous insulin infusion (CSII) is commonly used in patients with diabetes. Accurate and reliable delivery by insulin pumps is essential for a safe and effective therapy, particularly when using small doses. In this study, accuracy of bolus and basal rate delivery of various available insulin pumps was evaluated.
METHODS: In total, 13 insulin pump systems were tested: eight durable pumps with different infusion sets and one patch pump. Based on IEC 60601-2-24, insulin delivery was measured by recording weight gain of a beaker into which insulin was infused by the pumps. Bolus accuracy was determined by individually weighing 25 consecutive 0.1 or 1.0 U boluses and basal rate accuracy was determined during basal rate delivery of 0.1 or 1.0 U/h for 72 hours. For analyses, basal rate delivery was divided into 1-hour windows and deviation from target was calculated.
RESULTS: Regarding different systems, average 0.1 U bolus delivery was -2% to +9% of the intended volume with 53% to 96% of boluses within ±15% of target. During 0.1 U/h basal rate delivery, most pumps showed an initial over-delivery for the first few hours. Three systems reached a total basal rate error <5%; others showed up to +24%. In general, delivery was more accurate when using larger doses.
CONCLUSIONS: Considerable differences in insulin delivery accuracy were observed between the tested pumps. In general, when using very low doses, accuracy of insulin delivery is limited in most insulin pumps. This should be considered for CSII therapy in children.
© 2020 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  infusion pumps; insulin; insulin infusion systems

Mesh:

Substances:

Year:  2020        PMID: 32003490     DOI: 10.1111/pedi.12993

Source DB:  PubMed          Journal:  Pediatr Diabetes        ISSN: 1399-543X            Impact factor:   4.866


  2 in total

1.  Description of a Novel Patch Pump for Insulin Delivery and Comparative Accuracy Evaluation.

Authors:  Annette Baumstark; Jochen Mende; Joji Uchiyama; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2021-03-19

2.  Evaluation of the Accuracy and Reliability of a Tubeless Insulin Infusion System Under Laboratory Conditions.

Authors:  Jochen Mende; Manuel Eichenlaub; Delia Waldenmaier; Cornelia Haug; Guido Freckmann; Ralph Ziegler
Journal:  J Diabetes Sci Technol       Date:  2022-01-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.