Literature DB >> 31998545

Relationship between pin type and depth of skull penetration during frame placement for Gamma Knife radiosurgery.

Ronald E Warnick1, Eunsun Yook2.   

Abstract

PURPOSE: Our retrospective study addresses the largely unknown effect of fixation pin type on depth of skull penetration during Gamma Knife frame placement.
METHODS: Absolute and relative depths were compared for 404 pins of aluminum and titanium in 101 patients who underwent Gamma Knife frame placement at 0.4 Nm (3.5 lb-in) torque.
RESULTS: Effect of pin type was significant: penetration was less in aluminum than titanium pins, for absolute (0.27 mm vs. 0.83 mm) and relative (5.16% vs. 15.53%) depths (p < 0.001 each). Fifty percent of aluminum pins did not measurably penetrate the outer table of the skull whereas 95% of titanium pins penetrated into cancellous bone.
CONCLUSIONS: Although these pin penetration differences did not have a clear clinical correlate, we now recommend titanium pins for their consistent, deeper penetration, which may result in greater frame stability favorable for most clinical applications. Future studies may define optimal torque for aluminum pins.
© 2019 Old City Publishing, Inc.

Entities:  

Keywords:  Radiosurgery; complications; quality assurance; stereotaxic techniques

Year:  2019        PMID: 31998545      PMCID: PMC6774485     

Source DB:  PubMed          Journal:  J Radiosurg SBRT


  7 in total

Review 1.  The analysis of repeated measures designs: a review.

Authors:  H J Keselman; J Algina; R K Kowalchuk
Journal:  Br J Math Stat Psychol       Date:  2001-05       Impact factor: 3.380

2.  Stereotactic Shifts During Frame-Based Image-Guided Stereotactic Radiosurgery: Clinical Measurements.

Authors:  Sunil W Dutta; Roman O Kowalchuk; Daniel M Trifiletti; M Sean Peach; Jason P Sheehan; James M Larner; David Schlesinger
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-05-24       Impact factor: 7.038

3.  Relationships between head fixation pins for radiosurgery and the skull bone--usefulness of a torque wrench.

Authors:  Shun Toyota; Hidetoshi Seta; Hitoshi Kubo; Masatoshi Muramatsu; Kan Takeda
Journal:  Radiat Med       Date:  2003 Mar-Apr

4.  Complications of Head Immobilization Devices in Children: Contact Mechanics, and Analysis of a Single Institutional Experience.

Authors:  Mohamed A Zaazoue; Mostafa Bedewy; Liliana C Goumnerova
Journal:  Neurosurgery       Date:  2018-05-01       Impact factor: 4.654

5.  Techniques for the Application of Stereotactic Head Frames Based on a 25-Year Experience.

Authors:  Michael Safaee; John Burke; Michael W McDermott
Journal:  Cureus       Date:  2016-03-25

6.  Evaluation of the stability of the stereotactic Leksell Frame G in Gamma Knife radiosurgery.

Authors:  Alvaro Rojas-Villabona; Katherine Miszkiel; Neil Kitchen; Rolf Jäger; Ian Paddick
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

7.  Targeting inaccuracy caused by mechanical distortion of the Leksell stereotactic frame during fixation.

Authors:  Cécile Renier; Nicolas Massager
Journal:  J Appl Clin Med Phys       Date:  2019-04-04       Impact factor: 2.102

  7 in total
  1 in total

Review 1.  Role of endolysosome function in iron metabolism and brain carcinogenesis.

Authors:  Peter W Halcrow; Miranda L Lynch; Jonathan D Geiger; Joyce E Ohm
Journal:  Semin Cancer Biol       Date:  2021-06-15       Impact factor: 15.707

  1 in total