Literature DB >> 31997679

Accuracy of an electrogoniometer relative to optical motion tracking for quantifying wrist range of motion.

Brian P McHugh1, Amy M Morton2, Bardiya Akhbari1, Janine Molino1,2, Joseph J Crisco1,2.   

Abstract

Methods for capturing wrist range of motion (RoM) vary in complexity, cost, and sensitivity. Measures by manual goniometer, though an inexpensive modality, provide neither dynamic nor objective motion data. Conversely, optical motion capture systems are widely used in three-dimensional scientific motion capture studies but are complex and expensive. The electrogoniometer bridges the gap between portability and objective measurement. Our study aims to evaluate the accuracy of a 2 degree of freedom electrogoniometer using optical motion capture as the reference for in vivo wrist motion. First, a mechanical system constructed from two plastic pipes and a universal joint mimicked a human wrist to assess the inherent accuracy of the electrogoniometer. Simulations of radial/ulnar deviation (R/U), flexion/extension (F/E) and circumduction were evaluated. Second, six subjects performed three RoM tasks of R/U deviation, F/E, and circumduction for evaluation of the in vivo accuracy. Bland-Altman analysis quantified the accuracy. The mechanical experiment reported greater accuracy than the in vivo study with mean difference values less than ±1°. The in vivo accuracy varied across RoM tasks, with mean differences greatest in the F/E task (7.2°). Smaller mean differences values were reported in the R/U deviation task (-0.8°) and the circumduction task (1.2°).

Entities:  

Keywords:  Measurement error; goniometer; kinematics; motion analysis; wrist

Mesh:

Year:  2020        PMID: 31997679      PMCID: PMC7058510          DOI: 10.1080/03091902.2020.1713240

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  19 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  The effects of complex wrist and forearm posture on wrist range of motion.

Authors:  M M Marshall; J R Mozrall; J E Shealy
Journal:  Hum Factors       Date:  1999-06       Impact factor: 2.888

3.  A marker-based measurement procedure for unconstrained wrist and elbow motions.

Authors:  R Schmidt; C Disselhorst-Klug; J Silny; G Rau
Journal:  J Biomech       Date:  1999-06       Impact factor: 2.712

4.  In-vivo measurement of dynamic joint motion using high speed biplane radiography and CT: application to canine ACL deficiency.

Authors:  Scott Tashman; William Anderst
Journal:  J Biomech Eng       Date:  2003-04       Impact factor: 2.097

5.  Comparison of measurement accuracy between two wrist goniometer systems during pronation and supination.

Authors:  Peter W Johnson; Per Jonsson; Mats Hagberg
Journal:  J Electromyogr Kinesiol       Date:  2002-10       Impact factor: 2.368

6.  A new technique for the assessment of wrist movement using a biaxial flexible electrogoniometer.

Authors:  M L Rawes; J B Richardson; J J Dias
Journal:  J Hand Surg Br       Date:  1996-10

7.  Practical operation of a biaxial goniometer at the wrist joint.

Authors:  B Buchholz; H Wellman
Journal:  Hum Factors       Date:  1997-03       Impact factor: 2.888

8.  Kinematic Accuracy in Tracking Total Wrist Arthroplasty with Biplane Videoradiography using a CT-generated Model.

Authors:  Bardiya Akhbari; Amy Morton; Douglas Moore; Arnold-Peter C Weiss; Scott W Wolfe; Joseph Crisco
Journal:  J Biomech Eng       Date:  2019-02-07       Impact factor: 2.097

9.  Electrogoniometer measurement and directional analysis of wrist angles and movements during the Sollerman hand function test.

Authors:  Thomas Dauncey; Harvinder P Singh; Joseph J Dias
Journal:  J Hand Ther       Date:  2017-02-21       Impact factor: 1.950

10.  A novel validation and calibration method for motion capture systems based on micro-triangulation.

Authors:  Gergely Nagymáté; Tamás Tuchband; Rita M Kiss
Journal:  J Biomech       Date:  2018-04-12       Impact factor: 2.712

View more
  1 in total

Review 1.  Optical Motion Capture Systems for 3D Kinematic Analysis in Patients with Shoulder Disorders.

Authors:  Umile Giuseppe Longo; Sergio De Salvatore; Arianna Carnevale; Salvatore Maria Tecce; Benedetta Bandini; Alberto Lalli; Emiliano Schena; Vincenzo Denaro
Journal:  Int J Environ Res Public Health       Date:  2022-09-23       Impact factor: 4.614

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.