A new cardiac phantom for dynamic SPECT.

A Krakovich1, U Zaretsky2, I Moalem3, A Naimushin3, E Rozen3, M Scheinowitz2, R Goldkorn3.   

Abstract

BACKGROUND: In recent years, with the advance of myocardial blood flow (MBF) measurement capability in dynamic single photon emission computerized tomography (SPECT) systems, significant effort has been devoted to validation of the new capability. Unfortunately, the mechanical phantoms available for the validation process lack essential features-they either have a constant radiotracer concentration or they have rigid (static) walls unable to simulate cardiac beating. METHODS AND
RESULTS: We have developed a mechanical cardiac phantom that is able to mimic physiological radiotracer variation in the left ventricle (LV) cavity and in the myocardium (M), while performing beating-like motion. We have also developed a mathematical model of the phantom, allowing a description of the radiotracer concentrations in both regions (LV, M) as a function of time, which served as a tool for experiment planning and to accurately mimic physiological-like time-activity curves (TACs). A net retention model for the phantom was also developed, which served to compute the theoretical (i.e., expected) MBF of the phantom from measured quantities only, and thus validate the MBF reported by the SPECT system. In this paper, phantom experiments were performed on a GE Discovery NM 530c SPECT system.
CONCLUSIONS: A novel dynamic cardiac phantom for emission tomography has been developed. The new phantom is capable of producing a wide range of TACs that can mimic physiological (and potentially in the future, pathological) curves, similar to those observed in dynamic SPECT systems. SPECT-reported MBF values were validated against known (measured) activity of the injected radiotracer from phantom experiments, which allowed to determine the accuracy of the GE Discovery 530c SPECT system.
© 2020. American Society of Nuclear Cardiology.

Entities:  

Keywords:  CFR; MBF; Nuclear medicine; SPECT; cardiac phantom; dynamic phantom; heart; phantom

Mesh:

Substances:

Year:  2020        PMID: 31997101     DOI: 10.1007/s12350-020-02028-0

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  6 in total

1.  Absolute quantification of regional myocardial uptake of 99mTc-sestamibi with SPECT: experimental validation in a porcine model.

Authors:  A J Da Silva; H R Tang; K H Wong; M C Wu; M W Dae; B H Hasegawa
Journal:  J Nucl Med       Date:  2001-05       Impact factor: 10.057

2.  Validation of gated blood-pool SPECT cardiac measurements tested using a biventricular dynamic physical phantom.

Authors:  Pieter De Bondt; Kenneth Nichols; Stijn Vandenberghe; Patrick Segers; Olivier De Winter; Christophe Van de Wiele; Pascal Verdonck; Arsalan Shazad; Abu H Shoyeb; Johan De Sutter
Journal:  J Nucl Med       Date:  2003-06       Impact factor: 10.057

3.  Accuracy of 4 different algorithms for the analysis of tomographic radionuclide ventriculography using a physical, dynamic 4-chamber cardiac phantom.

Authors:  Pieter De Bondt; Tom Claessens; Bart Rys; Olivier De Winter; Stijn Vandenberghe; Patrick Segers; Pascal Verdonck; Rudi Andre Dierckx
Journal:  J Nucl Med       Date:  2005-01       Impact factor: 10.057

4.  Method of generating multiple sets of experimental phantom data.

Authors:  Arkadiusz Sitek; Bryan W Reutter; Ronald H Huesman; Grant T Gullberg
Journal:  J Nucl Med       Date:  2006-07       Impact factor: 10.057

5.  Optimization of SPECT Measurement of Myocardial Blood Flow with Corrections for Attenuation, Motion, and Blood Binding Compared with PET.

Authors:  R Glenn Wells; Brian Marvin; Marlie Poirier; Jennifer Renaud; Robert A deKemp; Terrence D Ruddy
Journal:  J Nucl Med       Date:  2017-06-13       Impact factor: 10.057

6.  Specific absorbed fractions for photon sources uniformly distributed in the heart chambers and heart wall of a heterogeneous phantom.

Authors:  J L Coffey; M Cristy; G G Warner
Journal:  J Nucl Med       Date:  1981-01       Impact factor: 10.057

  6 in total
  1 in total

1.  Dynamic phantoms: Making the right tool for the job.

Authors:  R Glenn Wells; Ran Klein
Journal:  J Nucl Cardiol       Date:  2020-03-02       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.