Evaluation of the Visibility and Artifacts of 11 Common Fiducial Markers for Image Guided Stereotactic Body Radiation Therapy in the Abdomen.

Jordan M Slagowski1, Lauren E Colbert2, Irina M Cazacu3, Ben S Singh3, Rachael Martin1, Eugene J Koay2, Cullen M Taniguchi2, Albert C Koong2, Manoop S Bhutani3, Joseph M Herman2, Sam Beddar4.   

Abstract

PURPOSE: The purpose of this study was to quantitatively evaluate the visibility and artifacts of commercially available fiducial markers to optimize their selection for image guided stereotactic body radiation therapy. METHODS AND MATERIALS: From 6 different vendors, we selected 11 fiducials commonly used in image guided radiation therapy. The fiducials varied in material composition (e.g., gold, platinum, carbon), shape (e.g., cylindrical, notched/linear, coiled, ball-like, step), and size measured in terms of diameter (0.28-1.0 mm) and length (3.0-20.0 mm). Each fiducial was centered in 4-mm bolus within a 13-cm-thick water-equivalent phantom. Fiducials were imaged with the use of a simulation computed tomography (CT) scanner, a CT-on-rails system, and an onboard cone beam CT system. Acquisition parameters were set according to clinical protocols. Visibility was assessed in terms of contrast (Δ Hounsfield unit [HU]) and the Michelson visibility metric. Artifacts were quantified in terms of relative standard deviation and relative streak artifacts level (rSAL). Twelve radiation oncologists ranked each fiducial in terms of clinical usefulness.
RESULTS: Contrast and artifacts increased with fiducial size. For CT imaging, maximum contrast (2722 HU) and artifacts (rSAL = 2.69) occurred for the largest-diameter (0.75 mm) platinum fiducial. Minimum contrast (551 HU) and reduced artifacts (rSAL = 0.65) were observed for the smallest-diameter (0.28 mm) gold fiducial. Carbon produced the least severe artifacts (rSAL = 0.29). The survey indicated that physicians preferred gold fiducials with a 0.35- to 0.43-mm diameter, 5- to 10-mm length, and coiled or cylindrical shape that balanced contrast and artifacts.
CONCLUSIONS: We evaluated 11 different fiducials in terms of visibility and artifacts. The results of this study may assist radiation oncologists who seek to maximize contrast, minimize artifacts, or balance contrast versus artifacts by fiducial selection.
Copyright © 2020 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2020        PMID: 31988039     DOI: 10.1016/j.prro.2020.01.007

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  7 in total

1.  A radiopaque polymer hydrogel as an irreversible electroporation compatible fiducial marker for pancreas stereotactic body radiotherapy.

Authors:  Eric M Anderson; Kenneth Park; Diana J Lu; Nicholas Nissen; Jun Gong; Andrew Hendifar; Alexandra Gangi; Simon Lo; Mitchell Kamrava; Katelyn M Atkins
Journal:  J Radiosurg SBRT       Date:  2020

Review 2.  Endoscopic ultrasound guided interventions in the management of pancreatic cancer.

Authors:  Tossapol Kerdsirichairat; Eun Ji Shin
Journal:  World J Gastrointest Endosc       Date:  2022-04-16

3.  Endoscopic Ultrasound-Guided Fiducial Placement for Stereotactic Body Radiation Therapy in Pancreatic Malignancy.

Authors:  Seong-Hun Kim; Eun Ji Shin
Journal:  Clin Endosc       Date:  2021-05-28

4.  Evaluation of a Novel Liquid Fiducial Marker, BioXmark®, for Small Animal Image-Guided Radiotherapy Applications.

Authors:  Kathryn H Brown; Mihaela Ghita; Giuseppe Schettino; Kevin M Prise; Karl T Butterworth
Journal:  Cancers (Basel)       Date:  2020-05-18       Impact factor: 6.639

5.  Experimental Comparison of Fiducial Markers Used in Proton Therapy: Study of Different Imaging Modalities and Proton Fluence Perturbations Measured With CMOS Pixel Sensors.

Authors:  Claire-Anne Reidel; Felix Horst; Christoph Schuy; Oliver Jäkel; Swantje Ecker; Katrin Henkner; Stephan Brons; Marco Durante; Uli Weber
Journal:  Front Oncol       Date:  2022-03-25       Impact factor: 6.244

6.  Precisely translating computed tomography diagnosis accuracy into therapeutic intervention by a carbon-iodine conjugated polymer.

Authors:  Mingming Yin; Xiaoming Liu; Ziqiao Lei; Yuting Gao; Jiacheng Liu; Sidan Tian; Zhiwen Liang; Ye Wang; Fanling Meng; Liang Luo
Journal:  Nat Commun       Date:  2022-05-12       Impact factor: 17.694

7.  EUS-guided placement of fiducial markers for image-guided radiotherapy in gastrointestinal tumors: A critical appraisal.

Authors:  Silvia Carrara; Mihai Rimbas; Alberto Larghi; Milena Di Leo; Tiziana Comito; Joseph Abi Jaoude; Cullen M Taniguchi; Christoph F Dietrich; Manoop S Bhutani; Stephan Hollerbach
Journal:  Endosc Ultrasound       Date:  2021 Nov-Dec       Impact factor: 5.628

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.