Literature DB >> 31985784

A novel multielectrode catheter for high-density ventricular mapping: electrogram characterization and utility for scar mapping.

Michael Barkagan1, Jakub Sroubek1, Ayelet Shapira-Daniels1, Hagai Yavin1, Jihye Jang2, Reza Nezafat2,3, Elad Anter1.   

Abstract

AIMS: Multielectrode mapping catheters can be advantageous for identifying surviving myocardial bundles in scar. This study aimed to evaluate the utility of a new multielectrode catheter with increased number of small and closely spaced electrodes for mapping ventricles with healed infarction. METHODS AND
RESULTS: In 12 swine (four healthy and eight with infarction), the left ventricle was mapped with investigational (OctarayTM) and standard (PentarayTM) multielectrode mapping catheters. The investigational catheter has more electrodes (48 vs. 20), each with a smaller surface area (0.9 vs. 2.0 mm2) and spacing is fixed at 2 mm (vs. 2-6-2 mm). Electrogram (EGM) characteristics, mapping efficiency and scar description were compared between the catheters and late gadolinium enhancement (LGE). Electrogram acquisition rate was faster with the investigational catheter (814 ± 126 vs. 148 ± 58 EGM/min, P = 0.02) resulting in higher density maps (38 ± 10.3 vs. 10.1 ± 10.4 EGM/cm2, P = 0.02). Bipolar voltage amplitude was similar between the catheters in normal and infarcted myocardium (P = 0.265 and P = 0.44) and the infarct surface area was similar between the catheters (P = 0.12) and corresponded to subendocardial LGE. The investigational catheter identified a higher proportion of near-field local abnormal ventricular activities within the low-voltage area (53 ± 16% vs. 34 ± 16%, P = 0.03) that were considered far-field EGMs by the standard catheter. The investigational catheter was also advantageous for mapping haemodymically non-tolerated ventricular tachycardias due to its higher acquisition rate (P < 0.001).
CONCLUSION: A novel multielectrode mapping catheter with higher number of small, and closely spaced electrodes increases the mapping speed, EGM density and the ability to recognize low amplitude near-field EGMs in ventricles with healed infarction. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author(s) 2020. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  Scar; Catheters; Electrodes; High-density mapping; Mapping; Substrate mapping; Ventricular tachycardia

Mesh:

Substances:

Year:  2020        PMID: 31985784     DOI: 10.1093/europace/euz364

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

1.  An early multicenter experience of the novel high-density star-shaped mapping catheter in complex arrhythmias.

Authors:  Andrea Sarkozy; Johan Vijgen; Tom De Potter; Richard Schilling; Vias Markides
Journal:  J Interv Card Electrophysiol       Date:  2022-03-26       Impact factor: 1.900

2.  Maximizing detection and optimal characterization of local abnormal ventricular activity in nonischemic cardiomyopathy: LAVAMAX & LAVAFLOW.

Authors:  Karl Magtibay; Stéphane Massé; Ahmed Niri; Robert D Anderson; Ram B Kumar; D Curtis Deno; Kumaraswamy Nanthakumar
Journal:  Heart Rhythm O2       Date:  2021-09-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.