Early Polyethylene Failure in a Modern Total Hip Prosthesis: A Note of Caution.

W Christian Thomas1, Hari K Parvataneni1, Richard G Vlasak1, Chancellor F Gray1.   

Abstract

BACKGROUND: Historically, there has been excellent survivorship of total hip arthroplasty with newer implants expected to outperform prior generations. Review of our institutional database identified problems with one particular cross-linked polyethylene liner in mid-term follow-up, which has not been reported in the literature. We report on the technical aspects of this liner, as well as the clinical and radiographic factors associated with early failure of this bearing in our practice.
METHODS: A retrospective review of our institutional database was performed from January 2009 to June 2019 of patients who presented with significant osteolysis in the setting of prior total hip arthroplasty with a contemporary polyethylene liner. Demographic and radiographic measures were collected for all patients.
RESULTS: Twelve patients were identified, with nine undergoing revision surgery. All patients demonstrated radiographic osteolysis at the time of revision surgery. The average time to diagnosis of failure was 55.9 months (range: 12-120 months). Average cup position preoperatively was 44 degrees inclination (range: 36-53 degrees) and 21 degrees anteversion (range: 10-34 degrees). Nine patients underwent revision of the acetabular component, with two also undergoing femoral component revision. The Manufacturer and User Facility Device Experience database revealed 22 reported cases of wear-related failure from 2009 to 2019.
CONCLUSIONS: Considering that no identifiable risk factors related to patient demographics or implant position were identified, the Exactech Connexion GXL liner may be prone to a high rate of early failure from wear and severe secondary osteolysis. We recommend close surveillance of patients with this bearing surface.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Exactech GXL; cross-linked polyethylene; mid-term failure; osteolysis; polyethylene wear

Year:  2020        PMID: 31982244     DOI: 10.1016/j.arth.2019.12.043

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

Review 1.  The lexicon for periprosthetic bone loss versus osteolysis after cervical disc arthroplasty: a systematic review.

Authors:  Jenna M Wahbeh; Sang-Hyun Park; Patricia Campbell; Edward Ebramzadeh; Sophia N Sangiorgio
Journal:  Eur Spine J       Date:  2022-01-09       Impact factor: 3.134

2.  Survival and performance of a dual tapered-wedge fully HA-coated press fit femoral stem.

Authors:  Ramakanth R Yakkanti; Dylan N Greif; Dennis J Vanden Berge; Raymond P Robinson
Journal:  Arch Orthop Trauma Surg       Date:  2022-03-03       Impact factor: 3.067

3.  Recent update on crosslinked polyethylene in total hip arthroplasty.

Authors:  Jean Langlois; Moussa Hamadouche
Journal:  SICOT J       Date:  2020-05-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.