Do Double-fan Surgical Helmet Systems Result in Less Gown-particle Contamination Than Single-fan Designs?

Alex Vermeiren1, Maarten Verheyden1, Frank Verheyden2.   

Abstract

BACKGROUND: Surgical helmet systems commonly are stand-alone systems with a single fan blowing air into the suit, creating positive pressure that blows particles out through areas of low resistance, possibly contaminating surgical attire and the surgical field. Two-fan systems were developed more recently to release spent air, also theoretically lowering pressure in the suit and decreasing the aforementioned risk of particle contamination. To our knowledge no study to date has measured the potential differences in gown particle contamination to support this hypothesis. QUESTIONS/PURPOSES: We compared a commonly used single-fan system versus a two-fan system and asked: (1) Which fan system results in less gown particle contamination? (2) Are there differences between the systems in the location of contamination?
METHODS: Using an existing experimental study model, two surgeons performed five 30-minute TKA simulations comparing a single-fan to a double-fan helmet system after applying fluorescent powder to the hands, axillae, and chest. Both are two-piece hood and gown systems. The single-fan sits on top of the helmet blowing air into the suit; the double-fan system has a second fan positioned at the rear blowing out spent air. Ultraviolet light-enhanced photographs were subsequently obtained of the flexor and extensor surfaces of the arms, axillary areas, and front and back of the chest. We chose these locations because they all contain either a seam or an overlap between gown and hood or gloves through which particles can escape. The images were scored for contamination on a scale of 1 (zero specks) to 4 (> 100 specks) by three independent observers. Interobserver correlation was assessed through Spearman's test yielding 0.91 (95% CI 0.86 to 0.94; p < 0.0001), 0.81 (95% CI 0.73 to 0.87; p < 0.0001) and 0.87 (95% CI 0.80 to 0.91; p < 0.0001) between observers 1 and 2, observers 1 and 3, and observers 2 and 3, rendering the used scale reliable. Results of the observers were averaged and compared using the Mann-Whitney U test.
RESULTS: There was no difference in overall gown particle contamination between the systems (overall single-fan median contamination score 2.5 of 4 [interquartile range Q1-Q3 0-3.42] versus double fan 1 out of 4 (Q1-Q3 0-3); p = 0.082), but all tests showed there was contamination at the gown-glove interface. In general, there were few differences between the two systems in terms of location of the contamination; however, when comparing only the axillary regions, we found that the single-fan group (median score 3.67 [Q1-Q3 3-4]) showed more contamination than the double-fan group (2.33 [Q1-Q3 0-3.08]); p = 0.01.
CONCLUSION: We found no difference in gown particle contamination between a single-fan and a double-fan helmet design. However, we note that contamination was present in all tests with both systems, so surgeons should not assume that these systems provide a contamination-free environment. CLINICAL RELEVANCE: When using such helmets, the surgeon should not place items close to the axillary region because the seam of the gown may have low resistance to particle contamination. Gown designs could be improved by creating better seals, especially at the arm-body seam.

Entities:  

Mesh:

Year:  2020        PMID: 31977431      PMCID: PMC7319402          DOI: 10.1097/CORR.0000000000001121

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.755


  14 in total

1.  Laminar air flow provides high air quality in the operating field even during real operating conditions, but personal protection seems to be necessary in operations with tissue combustion.

Authors:  Dorothea Hansen; Carsten Krabs; Daniel Benner; Andreas Brauksiepe; Walter Popp
Journal:  Int J Hyg Environ Health       Date:  2005-10-06       Impact factor: 5.840

2.  The Gown-glove Interface Is a Source of Contamination: A Comparative Study.

Authors:  James F Fraser; Simon W Young; Kimberly A Valentine; Nicholas E Probst; Mark J Spangehl
Journal:  Clin Orthop Relat Res       Date:  2015-07       Impact factor: 4.176

3.  Intraoperative contamination and space suits: a potential mechanism.

Authors:  Simon W Young; Carl Chisholm; Mark Zhu
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-02-09

4.  Comparison of air exhausts for surgical body suits (space suits) and the potential for periprosthetic joint infection.

Authors:  F Ling; S Halabi; C Jones
Journal:  J Hosp Infect       Date:  2018-03-17       Impact factor: 3.926

5.  A novel total knee arthroplasty infection model in rabbits.

Authors:  Matthew R Craig; Kornelis A Poelstra; J Christopher Sherrell; Michael S Kwon; Etienne L Belzile; Thomas E Brown
Journal:  J Orthop Res       Date:  2005-09       Impact factor: 3.494

6.  Risk factors associated with surgical site infection in 30,491 primary total hip replacements.

Authors:  R S Namba; M C S Inacio; E W Paxton
Journal:  J Bone Joint Surg Br       Date:  2012-10

7.  The Impact of Patient and Surgical Factors on the Rate of Postoperative Infection After Total Hip Arthroplasty-A New Zealand Joint Registry Study.

Authors:  James O Smith; Christopher M A Frampton; Gary J Hooper; Simon W Young
Journal:  J Arthroplasty       Date:  2018-01-31       Impact factor: 4.757

8.  The influence of the total body exhaust suit on air and wound contamination in elective hip-operations.

Authors:  G Blomgren; A Hambraeus; A S Malmborg
Journal:  J Hosp Infect       Date:  1983-09       Impact factor: 3.926

9.  Lack of influence of body exhaust gowns on aerobic bacterial surface counts in a mixed-ventilation operating theatre. A study of 62 hip arthroplasties.

Authors:  C Pasquarella; O Pitzurra; T Herren; L Poletti; A Savino
Journal:  J Hosp Infect       Date:  2003-05       Impact factor: 3.926

10.  Effect of ultraclean air in operating rooms on deep sepsis in the joint after total hip or knee replacement: a randomised study.

Authors:  O M Lidwell; E J Lowbury; W Whyte; R Blowers; S J Stanley; D Lowe
Journal:  Br Med J (Clin Res Ed)       Date:  1982-07-03
View more
  1 in total

1.  CORR Insights®: Do Double-fan Surgical Helmet Systems Result in Less Gown-particle Contamination Than Single-fan Designs?

Authors:  Simon W Young
Journal:  Clin Orthop Relat Res       Date:  2020-06       Impact factor: 4.755

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.