| Literature DB >> 31974281 |
Christopher Yusuf Akhunbay-Fudge1,2, Kenan Deniz1, Atul Kumar Tyagi1, Tufail Patankar3.
Abstract
BACKGROUND ANDEntities:
Keywords: aneurysm; angiography; device
Mesh:
Year: 2020 PMID: 31974281 PMCID: PMC7509519 DOI: 10.1136/neurintsurg-2019-015628
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1The Cerus Endovascular Contour Neurovascular System (A–C) including angiographic view (D) showing detachment zone (arrow).
Figure 2(A) Target zone of Contour Neurovascular System (CNS) device deployment at the intracranial aneurysm neck and within the equatorial plane. (B) Schematic representation of the CNS device with catheter attached. (C) Schematic representation of the CNS device deployed in the aneurysm showing expected flow diversion. (D) Schematic representation of angiographic view of the CNS device.
Sizing guide for the Contour Neurovascular System
| Catalogue number | Aneurysm neck (mm) | Aneurysm width |
| CNS05 −5 mm | 2.0–3.0 | 2.0–3.5 |
| CNS07 −7 mm | 3.0–5.0 | 3.0–5.5 |
| CNS09 −9 mm | 4.0–6.0 | 5.0–7.5 |
| CNS11 –11 mm | 5.0–8.0 | 7.0–8.5 |
Summary of clinical and radiological results of Contour Neurovascular System series
| ID | Lesion location | Aneurysm dimensions | AR | DNR | Device | Raymond–Roy class | ||||
| Neck (mm) | Width (mm) | Height (mm) | Month 6 | Month 12 | Month 24 | |||||
| 01 | ICA | 2.4 | 3.5 | 4.3 | 1.8 | 1.5 | 7 | Class 1 | Class 1 | Class 1 |
| 02 | ICA | 3.9 | 8.7 | 9.1 | 2.3 | 2.2 | 9 | Class 2 | Class 2 | Class 2 |
| 03 | Basilar | 3.2 | 6.7 | 7.8 | 2.4 | 2.1 | 11 | Class 1 | Class 1 | Class 1 |
| 04 | SCA | 3.6 | 5.1 | 6.7 | 1.9 | 1.4 | 9 | Class 2 | Class 2 | Class 2 |
| 05 | Basilar | 2.7 | 4.2 | 5.5 | 2.0 | 1.6 | 9 | Class 1 | Class 1 | Class 1 |
| 06 | Basilar | 3.5 | 5.9 | 6.5 | 1.9 | 1.7 | 11 | Class 3 | n/a | n/a |
| 07 | ACOM | 3.2 | 4.3 | 5.8 | 1.8 | 1.3 | 7 | Class 1 | Class 1 | n/a |
| 08 | MCA | 5.4 | 10.2 | 12.0 | 2.2 | 1.9 | 11 | Class 3 | Class 1 | n/a |
| 09 | MCA | 4.4 | 7.5 | 13.0 | 3.0 | 1.7 | 11 | Class 3 | Class 2 | n/a |
| 10 | MCA | 4.8 | 6.5 | 7.1 | 1.5 | 1.4 | 11 | Class 2 | n/a* | n/a |
| 11 | Basilar | 4.0 | 3.8 | 6.1 | 1.5 | 1.0 | 7 | Class 2 | Class 2 | Class 2 |
*Occluded on 1 year MR angiography.
ACOM, anterior communicating artery; AR, aspect ratio (height/neck); DNR, dome/neck ratio (width/neck)); ICA, internal carotid artery; MCA, middle cerebral artery; SCA, superior cerebellar artery.
Summary of adverse events
| ID | Thromboembolic events | Intraprocedural rupture | Morbidity | Mortality | ||
| A | SS | SWS | ||||
| 01 | No | No | No | No | No | No |
| 02 | No | No | No | No | No | No |
| 03 | No | No | No | No | Headaches | No |
| 04 | No | No | No | No | No | No |
| 05 | No | No | Yes | No | No | No |
| 06 | No | No | No | No | No | No |
| 07 | No | No | No | No | No | No |
| 08 | No | No | No | No | No | No |
| 09 | No | No | No | No | No | No |
| 10 | No | No | Yes | No | Headaches | No |
| 11 | No | No | No | No | No | No |
A, asymptomatic; SS, symptomatic with sequelae; SWS, symptomatic without sequelae.
Summary of failed cases with the Contour Neurovascular System
| ID | Lesion location | Aneurysm dimensions | AR | DNR | Procedural difficulties | ||
| Neck (mm) | Width (mm) | Height (mm) | |||||
| 12 | MCA | 3.3 | 4.6 | 4.4 | 1.3 | 1.4 | Unable to catheterize with XT 27 catheter so coils were attempted. Despite eventual catheterization with Echelon 10 microcatheter, balloon inflation and attempts with multiple coils, the procedure was felt to be high risk of stroke and abandoned |
| 13 | Pericallosal | 3.7 | 6.6 | 5.5 | 1.5 | 1.8 | Inability to catheterize aneurysm with XT 27 catheter therefore successfully reverted to coiling |
| 14 | MCA | 5.1 | 6.3 | 7.2 | 1.4 | 1.2 | An XT 27 was navigated into the aneurysm. There were two failed trials with two 11 mm Contour devices. Subsequently a 9×4 mm WEB was used but was not suitable and an 8×3 mm WEB was used instead with good clinical and radiological outcomes |
AR, aspect ratio (height/neck); DNR, dome/neck ratio (width/neck); MCA, middle cerebral artery.
Figure 3Case 1. (A) Three-dimensional reconstruction of the internal carotid artery aneurysm. (B) Angiographic view of untreated aneurysm. (C) Microcatheter at aneurysm neck. (D) 7 mm Contour Neurovascular System deployed within aneurysm. (E) Early angiographic view showing signs of stasis. (F) Six-month follow-up MRI showing complete occlusion, although with artefact. (G) Six-month follow-up angiogram showing complete occlusion and flow diversion. (H) Angiographic view showing device remains in good position.