Reflotron cholesterol measurement evaluated as a screening technique.

F A Sedor1, C M Holleman, S Heyden, K A Schneider.   

Abstract

We evaluated the analytical performance of Boehringer Mannheim Diagnostics' "Reflotron" analyzer for the measurement of cholesterol. Coefficients of variation (CVs) for whole-blood cholesterol were: within-day 2.0% and 2.2% at 1680 and 2670 mg/L, respectively; between-day 1.8% and 2.4% (n = 9 and 8). Results were similar for serum and heparinized or EDTA-treated single-donor plasma (CV 1.4% to 2.6%). CVs of results for two reconstituted commercial quality-control materials were 3.4% and 4.6%. Heparin and hematocrit were evaluated as interferents, and critical limits for interference were identified for bilirubin, hemoglobin, and triglyceride in blood and plasma or serum. When sample collection and analysis were controlled by trained personnel, results with the Reflotron (y) compared well with those by the Ektachem procedure (x) for both blood and serum samples: r = 0.950, y = 0.944x + 130 mg/L; and r = 0.955, y = 0.93x + 43.5 mg/L, respectively. The same comparability was observed when the analysis was performed by briefly trained high-school students: r = 0.980, y = 0.949x + 23 mg/L. Performance decreased when both collection and analysis were performed by laymen: r = 0.880, y = 0.870x + 186 mg/L.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3197298

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  Near-patient testing in primary care.

Authors:  David Fitzmaurice
Journal:  Br J Gen Pract       Date:  2004-09       Impact factor: 5.386

2.  Performance of the Reflotron in Massachusetts' Model System for Blood Cholesterol Screening Program.

Authors:  S Havas; R Bishop; L Koumjian; J Reisman; S Wozenski
Journal:  Am J Public Health       Date:  1992-03       Impact factor: 9.308

3.  Measurement of capillary cholesterol in hyperlipidemia.

Authors:  J L Curzio; C Howie; S Kennedy; J L Reid; E Farish; J Barnes
Journal:  Br J Gen Pract       Date:  1991-10       Impact factor: 5.386

4.  Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.

Authors:  G Thue; S Sandberg; D G Bullock
Journal:  Br J Gen Pract       Date:  1993-01       Impact factor: 5.386

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.