Evaluation of a Novel Spun Polytetrafluoroethylene Stent Graft in an Ovine External Iliac Artery Model.

Bart L Dolmatch1, John W Hall2, Wayne L Mower3, Serge D Rousselle4.   

Abstract

PURPOSE: To evaluate the patency, cellular response, and thrombogenicity of a novel vascular stent graft.
MATERIALS AND METHODS: Test stent grafts, incorporating luminal spun polytetrafluoroethylene and a nonpermeable fluoropolymer layer, and control stent grafts, constructed of permeable expanded polytetrafluoroethylene, were implanted in the external iliac arteries of 14 adult sheep with a median weight of 73.4 kg ranging from 60.6-86.8 kg for 30 (n = 4), 90 (n = 4), and 180 (n = 6) days. Angiographic patency and percent diameter stenosis (%DS) were assessed at termination. Excised stent grafts were fixed and stained for histopathologic analysis, including neointimal coverage (NC) assessment.
RESULTS: Test and control device migration occurred in 1 animal, resulting in test device thrombosis. Both devices were excluded from analysis. Mean %DS in test and control implants was 4.6% and 8.2% (P = .563), 2.0% and 10.9% (P = .363), and 2.1% and 10.3% (P = .009) at 30, 90, and 180 days, respectively. Median NC scores at 30, 90, and 180 days were significantly lower in middle test device sections (P < .05). Proximal and distal test and control sections exhibited similar median NC scores at all time periods (P > .05). When present, test and control devices exhibited no neointimal detachment from the graft surface. Except for the migrated test device, no thrombus was observed. Transgraft cellular migration was absent in test devices but present in control devices with tissue accumulation around the stent struts.
CONCLUSIONS: Test and control devices demonstrated excellent patency in an ovine model. Compared to the control, test devices exhibited significantly lower %DS values at 180 days and significantly lower mid-device NC scores at 30, 90, and 180 days.
Copyright © 2019 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2020        PMID: 31959517     DOI: 10.1016/j.jvir.2019.07.036

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  1 in total

1.  First Clinical Results of the Merit WRAPSODY™ Cell-Impermeable Endoprosthesis for Treatment of Access Circuit Stenosis in Haemodialysis Patients.

Authors:  James Gilbert; Jason Rai; David Kingsmore; John Skousen; Nikolaos Ptohis
Journal:  Cardiovasc Intervent Radiol       Date:  2021-09-12       Impact factor: 2.740

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.