Hui Wang1,2,3, Wenjie Wu2,3,4, Michael Talcott5, Robert C McKinstry6, Pamela K Woodard4,6, George A Macones7, Alan L Schwartz8, Phillip Cuculich9, Alison G Cahill7, Yong Wang2,3,4,6.
Abstract
The dataset presented in this paper is related to the recent work "Accuracy of electromyometrial imaging of uterine contractions in clinical environment" [1]. The dataset including body-uterus geometry obtained from magnetic resonance imaging (MRI), uterine electrograms and isochrone maps reconstructed using Electromyometrial imaging (EMMI) under various levels of deformations and electrical noise contamination in a translational sheep model are reported. The dataset make it possible for detailed evaluation and further improvement of EMMI. In addition, the researchers working on other types of electrophysiology imaging techniques, such as electrocardiographic imaging (ECGI), and Electrogastrography imaging (EGGI) could also adopt our method [1] and employ the dataset to evaluate and improve their imaging techniques.Entities:
Keywords: Electrical noise; Electromyograms; Electromyometrial imaging; Geometry; Isochrone maps; Sheep model
Year: 2020 PMID: 31956675 PMCID: PMC6956746 DOI: 10.1016/j.dib.2019.105078
Source DB: PubMed Journal: Data Brief ISSN: 2352-3409
Fig. 1Body-uterus geometry. a, Sheep body-uterus geometry with unipolar electrode locations, red dots represent body surface electrode locations and blue dots represent uterine surface electrode locations. b, Discretized body-uterus geometry implemented in the analysis [1]. c, Full uterine surface recorded from MRI (pink) and uterine surface electrode locations (Blue dots).
Fig. 2Gaussian and Perlin noise. a, b Simulated noise electrogram of the first 10 seconds of one electrode at SNR = 17.5 dB. c, d noise amplitude per Hz of electrograms in a, b. The total noise power is mW in a and 7.35 mW in b. The total noise power under 10 Hz is mW in a and mW in b.
Fig. 3Type I deforming process and extreme deformation ratio distribution. a, Type I deforming processing. As for Laplacian deformation algorithm, a chosen handle was moved to represent the kick center (blue sphere), a kick direction was defined as outwards from the center of the uterus (red arrow), and a non-stationary area was defined as deformed (the area from yellow belt to the center). b, c local and global deformation, respectively. In each row, the four images show deformation ratios in heat maps from the four indicated views. Warm colors indicate large deformations, and cool colors indicate small deformations as indicated in scales at far right.
Fig. 4Electrograms under extreme deformations or noise. a, Four views of sheep uterus; light pink is the uterine geometry segmented from MRI images; dark pink is the discretized uterine geometry from the locations of electrodes placed on the uterine surface. b, Electrograms reconstructed from original body surface potentials from four locations labeled A, B, C, and D in a. c–h, Reconstructed electrograms corresponding to the deformed or noise-contaminated body surface potentials. Deformation extents are at the extreme conditions, and SNR = 12.5 dB.
Fig. 5EMMI-reconstructed isochrone maps during another EMG burst. In the top seven rows, isochrones are displayed for two deformation or noise levels and shown in four views. In the heat maps, red indicates early activation, blue indicates late activation, and the darkest blue denotes regions in which no activation was recorded during the specific observation window. The black star indicates the location at which the electrogram (bottom row) was reconstructed. The electrogram shows EMG bursts (denoted by the red step lines); blue and green vertical lines indicate the start and end of the observation window for EMG burst (from 286 s to 378 s). a - d, Isochrones from original, local deformed, global deformed, and noise-contaminated electrograms, respectively. Black circles in panel b mark local deformation centers.
Specifications Table
| Subject | Biomedical Engineering |
| Specific subject area | Electrophysiology; Imaging technique |
| Type of data | Figures |
| How data were acquired | Body-uterus geometry was imaged by Magnetic resonance imaging (MRI) and segmented using Amira software (Version 6.2). Electrograms and isochrone maps under deformation or electrical noise were reconstructed by Electromyometrial imaging (EMMI). |
| Data format | Raw computational |
| Parameters for data collection | Images of uterus and abdomen were acquired using human MRI scanner from an anesthetized sheep. The sheep uterine electrograms were recorded during a surgical procedure to expose the sheep uterus [ |
| Description of data collection | MRI images of uterus and abdomen were acquired using a radial VIBE fast T1-weighted sequence from an anesthetized sheep. MRI images were segmented and rendered to uterus and body mesh. The sheep uterus was exposed surgically, and an elastic sock containing 64 or 128 sintered Ag–AgCl electrodes was slipped over the uterus to record the electrograms at a 2 kHz sampling rate using Biosemi system [ |
| Data source location | Washington University in St. Louis School of Medicine, St. Louis, Missouri, United States |
| Data accessibility | The data are available with the article. |
| Related research article | H. Wang, W. Wu, M. Talcott, S. Lai, R.C. McKinstry, P.K. Woodard, G.A. Macones, A.L. Schwartz, P.S. Cuculich, A.G. Cahill, Y. Wang, Accuracy of electromyometrial imaging of uterine contractions in clinical environment, |
The computational and processed data will enable readers to reproduce all our findings [ The data and simulation of geometrical changes can also be adapted to assess the accuracy of other types of electrophysiology imaging methods, such as electrocardiographic imaging (ECGI) [ Readers and researchers can take our work as the base and continue to incorporate more realistic geometrical deformation models and electrical noises to generate insights about the clinical EMMI translation and data interpretation. |