Ultrasound-guided non-targeted liver core biopsy: comparison of the efficacy of two different core needle biopsy systems using an ex-vivo animal model and retrospective review of clinical experience.

Lisa M Ho1, Avani A Pendse2, James Ronald3, Marcus Luciano4, Daniele Marin5, Tracy A Jaffe6, Rendon C Nelson7.   

Abstract

PURPOSE: To compare the efficacy of two 18-gauge core needle biopsy systems, the Achieve® (Merit Medical) and the Marquee® (BD Bard), using an ex-vivo animal liver model and retrospective review of clinical experience.
METHODS: Sixty ex-vivo liver biopsy samples were obtained using the Achieve® (n = 30) and the Marquee® (n = 30) needles. In addition, 20 liver biopsy samples from 20 patients obtained using the Achieve® (n = 10) and Marquee® (n = 10) were compared retrospectively. One pathologist, blinded to needle type, recorded total core length and the number of complete portal triads. Ex vivo measurements were compared using mixed effects linear, logistic, and ordinal regression. In vivo measurements were compared using Student's t-test.
RESULTS: For the Achieve® and Marquee® needles, the mean(SD) total core length (mm) of ex vivo samples was 11.0(3.3) and 12.6(3.4), respectively (P = 0.069) and the adequacy rate was 23.3% and 50%, respectively (P = 0.04). Mean number of portal triads of ex vivo samples was 7.2(2.9) and 8.6(3.8), respectively (P = 0.13), and the adequacy rate was 73.3% and 83.3%, respectively (P = 0.32). For in vivo samples, the Achieve® and Marquee® needles demonstrates mean(SD) total core length (mm) of 24.6(7.1) and 32.0(4.6), respectively (P = 0.01), adequacy rate (P = 0.06). Mean number of portal triads was 14.9(4.8) and 19.6(4.1), respectively (P = 0.03), adequacy rate (P = 0.47).
CONCLUSIONS: Slightly longer core biopsies were obtained with the Marquee® needle compared with the Achieve® needle. Early clinical experience demonstrates no significant difference in sample adequacy rates. Both needle types can be expected to provide adequate samples for pathologic assessment of liver tissue.
Copyright © 2020. Published by Elsevier Inc.

Entities:  

Year:  2020        PMID: 31954350     DOI: 10.1016/j.clinimag.2020.01.005

Source DB:  PubMed          Journal:  Clin Imaging        ISSN: 0899-7071            Impact factor:   1.605


  1 in total

1.  Needle types used in abdominal cross-sectional interventional radiology: a survey of the Society of Abdominal Radiology emerging technology commission.

Authors:  Benjamin Wildman-Tobriner; Lisa M Ho; Andrew W Bowman
Journal:  Abdom Radiol (NY)       Date:  2021-06-14
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.