Effect of Different ECG Leads on Estimated R-R Intervals and Heart Rate Variability Parameters.

Vala Jeyhani, Matti Mantysalo, Kai Noponen, Tapio Seppanen, Antti Vehkaoja.   

Abstract

Heart rate and heart rate variability parameters provide important information on sympathetic and parasympathetic branches of autonomous nervous system. These parameters are usually extracted from electrocardiograms often measured between two electrodes and called an ECG lead. Besides systems intended only for heart rate measurement, ECG measurement devices employ several well-known lead systems including the standard 12-lead system, EASI lead system and Mason-Likar systems. Therefore, the first step is to select the appropriate lead for heart rate variability analysis. The appropriate electrode locations for single-lead measurement systems or the preferred measurement lead in multi-lead measurement are choices that the user needs to make when the heart rate variability is of interest. However, it has not been addressed in the literature, if the lead selection has an effect on the obtained HRV parameters. In this work, we characterized the amount of deviation of heart rate and heart rate variability parameters extracted from nine ECG leads, six from EASI leads and three modified limb leads. The results showed a deviation of 2.04, 2.88, 2.06 and 3.45 ms in SDNN, rMSSD, SD1 and SD2, respectively. A relative difference up to 10% was observed in HRV parameters for single signal frames. Additionally, the discrimination of the R-peaks by amplitudes was evaluated. The A-S lead appeared to have the best performance in all the tests.

Entities:  

Mesh:

Year:  2019        PMID: 31946698     DOI: 10.1109/EMBC.2019.8857954

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  5 in total

1.  Heart Rate Variability in Healthy Subjects During Monitored, Short-Term Stress Followed by 24-hour Cardiac Monitoring.

Authors:  Zifan Gu; Vanessa C Zarubin; Katherine R Mickley Steinmetz; Carolyn Martsberger
Journal:  Front Physiol       Date:  2022-06-13       Impact factor: 4.755

Review 2.  Fractal Correlation Properties of Heart Rate Variability as a Biomarker for Intensity Distribution and Training Prescription in Endurance Exercise: An Update.

Authors:  Bruce Rogers; Thomas Gronwald
Journal:  Front Physiol       Date:  2022-05-09       Impact factor: 4.755

3.  Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women.

Authors:  Marcelle Schaffarczyk; Bruce Rogers; Rüdiger Reer; Thomas Gronwald
Journal:  Sensors (Basel)       Date:  2022-08-30       Impact factor: 3.847

4.  Estimation of Respiratory Frequency in Women and Men by Kubios HRV Software Using the Polar H10 or Movesense Medical ECG Sensor during an Exercise Ramp.

Authors:  Bruce Rogers; Marcelle Schaffarczyk; Thomas Gronwald
Journal:  Sensors (Basel)       Date:  2022-09-21       Impact factor: 3.847

5.  The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study.

Authors:  Bruce Rogers; Marcelle Schaffarczyk; Martina Clauß; Laurent Mourot; Thomas Gronwald
Journal:  Sensors (Basel)       Date:  2022-03-05       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.