Literature DB >> 31946410

Validation of Noninvasive Ultrasound Temperature Measurement System Through Experiments Using Resonant Cavity Applicator.

Yasuhiro Shindo, Kazuo Kato, Futoshi Ikuta, Kenji Takahashi.   

Abstract

We developed a resonant cavity applicator for noninvasive deep hyperthermia treatment of osteoarthritis. In previous studies, we confirmed its viability by heating agar phantoms and conducting computer simulations. To evaluate the heating performance of this system on human subjects, it is necessary to develop a new noninvasive temperature measurement system. In this study, we developed a new temperature measurement system utilizing ultrasound imaging techniques. After heating an agar phantom with our resonant cavity applicator, temperature measurements from both our custom-made noninvasive ultrasonic thermometer and an optical fiber thermometer were collected and compared. Both temperature distributions showed a reliable trend, with heat energy concentrated at the center of the agar phantom. Average error rates were less than 13% for temperatures. The results of this study describe the viability of the temperature measurement system and the resonant cavity applicator for clinical applications.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 31946410     DOI: 10.1109/EMBC.2019.8856285