A PHANTOM STUDY SHOWING THE IMPORTANCE OF BREAST SHIELDING DURING HEAD CT.

Nika Zalokar1, Nejc Mekiš1.   

Abstract

This study aimed to investigate the dose to the breasts during head computed tomography (CT) if lead shielding is used. The study was performed in two major hospitals using helical and axial protocols on an anthropomorphic phantom. Measurements were performed with and without the use of a lead shield of 0.5 mm equivalent density. The results showed a significant decrease in dose with the lead shielding in both hospitals. During the helical protocol, the use of shielding significantly reduced the dose by 96% in Hospital A and 82% in Hospital B. The dose reduction during axial protocol was also significant: 95% in Hospital A and 86% in Hospital B with lead shielding. Considering the significant dose reduction of 82% up to 96% during this study, we highly recommend the shielding of breasts regardless of the protocol used during head CT examinations.
© The Author(s) 2020. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 31942634     DOI: 10.1093/rpd/ncz305

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  2 in total

1.  Efficacy of breast shielding during head computed tomography examination.

Authors:  Nika Zalokar; Nejc Mekis
Journal:  Radiol Oncol       Date:  2020-08-01       Impact factor: 2.991

2.  Efficacy of breast shielding during head computed tomography examination.

Authors:  Nika Zalokar; Nejc Mekis
Journal:  Radiol Oncol       Date:  2020-08-01       Impact factor: 2.991

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.