Tsokyi Choera1, Brittney Jung-Hynes2, Derrick J Chen3.
Abstract
BACKGROUND: Group B Streptococcal (GBS) infections in the United States are a leading cause of meningitis and sepsis in newborns. The CDC therefore recommends GBS screening for all pregnant women at 35-37 weeks of gestation and administration of intrapartum prophylaxis (in those that tested positive) as an effective means of controlling disease transmission. Several FDA approved molecular diagnostic tests are available for rapid and accurate detection of GBS in antepartum women.Entities:
Keywords: GBS comparison; Group B Streptococcus; Molecular test; Nucleic acid amplification test; Perinatal prevention; Streptococcus agalactiae
Mesh:
Substances:
Year: 2020 PMID: 31937247 PMCID: PMC6958782 DOI: 10.1186/s12879-019-4756-y
Source DB: PubMed Journal: BMC Infect Dis ISSN: 1471-2334 Impact factor: 3.090
Comparison of results for the Xpert GBS LB Assay, the Revogene GBS LB Assay, and Culture
| No. of specimens | Test Results | ||||
|---|---|---|---|---|---|
| Xpert GBS LB assay | Revogene GBS LB assay | Repeat Xpert | Repeat Revogene | Culture* | |
| 192 | Negative | Negative | N/A | N/A | N/A |
| 50 | Positive | Positive | N/A | N/A | N/A |
| 3 | Negative | Positive | Positive | Positive | Negative |
| 2 | Negative | Positive | Negative | Negative | Negative |
| 1 | Positive | Negative | Negative | Negative | Negative |
| 1 | Negative | Positive | Negative | Positive | Negative |
| 1 | Negative | Positive | Negative | Positive | Positive |
*Culture was only performed for discrepant results between Xpert and Revogene
Performance characteristics of Revogene compared to Xpert GBS LB Assays
| Xpert | ||||
|---|---|---|---|---|
| Positive | Negative | Total | ||
| Revogene | Positive | 50 | 7 | 57 |
| Negative | 1 | 192 | 193 | |
| Total | 51 | 199 | 250 | |
| 95% CI | ||||
| Positive Percent Agreement | 98.0% | 89.6 to 99.7% | ||
| Negative Percent Agreement | 96.5% | 92.9 to 98.3% | ||
| Overall agreement | 96.8% | 93.8 to 98.4% | ||
Time and Cost Comparison between the Xpert GBS LB and the Revogene GBS LB Assays
| Assay | Time (mins) | Cost ($)a | ||
|---|---|---|---|---|
| Setup Per Test | Run Time | Per Testa | Per Instrumenta | |
| Xpert GBS LB | 1.3 | 59 | 30 | 110,000 |
| Revogene GBS LB | 2.7 | 70 | 28 | 35,000 |
aCost is based on the list price provided by the vendors; each vendor works with individual labs to try to make these kits and instrument affordable within their budgets
**The test price is listed as per test; however, the tests are purchased as kits