As the techniques to connect percutaneous coronary intervention (PCI) balloons and the inflation syringe vary in the instructions for use and in practice, we measured the amount of air in PCI balloons after testing three connection methods to an inflation syringe. Following the preparation using one of the three methods, 114 balloons and stent balloons were tested four times. Method 1 connected the syringe and the balloon catheter directly after purging and filling the lumen, while method 3 omitted the purging and filling process. With method 2, the catheter lumen was purged, filled and fully vented via a three-way valve. The primary endpoint answered whether air remained in the balloon, and if so, the secondary endpoint indicated the total volume of remaining air. The connection with a three-way valve achieved significantly less air in the inflated balloon as compared with either direct connection approach (27% vs. 44% and 51%; p = 0.015). For the direct connection, no significant difference between purging and filling the lumen prior to making the connection or not existed. According to these findings, the best method to connect a PCI balloon to the inflation syringe while removing air involves using a three-way valve.Entities:
Keywords: acute coronary syndrome (ACS); air remocal; balloon inflation; balloon preparation; percutaneous coronary intervention (PCI)
Year: 2020 PMID: 31936381 PMCID: PMC7020097 DOI: 10.3390/jcm9010172
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Groups of balloons and stents.
| Diameter | Length | ||
|---|---|---|---|
|
| Group 1 | ≤2.5 mm | ≤18.0 mm |
| Group 2 | ≤2.5 mm | ≥20.0 mm | |
| Group 3 | ≥3.0 mm | ≤18.0 mm | |
| Group 4 | ≥3.0 mm | ≥20.0 mm | |
Figure 1Methods of connecting the balloons.
Figure 2Balloon without air.
Figure 3Balloon containing air.
Figure 4Balloon with two air bubbles.
Total volume categories.
| Volume in mm3 | ||
|---|---|---|
|
| High | ≥1.35 |
| Medium | ≥0.51, <1.35 | |
| Low | <0.51 | |
Figure 5Excluded balloons group 1.
Figure 6Excluded balloons group 2.
Figure 7Excluded balloons group 3.
Figure 8Excluded balloons group 4.
Air in the balloons depending on the method.
| Air | |||
|---|---|---|---|
| Yes | No | ||
|
| 1 (direct connection with filling the lumen) | 44% | 56% |
| 2 (three-way valve) | 27% | 73% | |
| 3 (direct connection without filling the lumen) | 51% | 49% | |
|
| 41% | 59% | |
Pairwise comparison of the methods.
| 95% Confidence Interval | |||||
|---|---|---|---|---|---|
| (I) Method | (J) Method | Mean Difference (I-J) | Significance | Lower Value | Upper Value |
| 1 (direct connection with filling the lumen) | 2 | −0.17 | 0.028 | −0.33 | −0.02 |
| 3 | 0.08 | 0.320 | −0.08 | 0.23 | |
| 2 (three-way valve) | 1 | 0.17 | 0.028 | 0.02 | 0.33 |
| 3 | 0.25 | 0.002 | 0.09 | 0.41 | |
| 3 (direct connection without filling the lumen) | 1 | −0.08 | 0.320 | −0.23 | 0.08 |
| 2 | −0.25 | 0.002 | −0.41 | −0.09 | |
Air in the balloons for the different groups.
| Air | |||
|---|---|---|---|
| Group | Method | Yes | No |
| 1 (short, small diameter) | 1 (direct connection with filling the lumen) | 46% | 54% |
| 2 (three-way valve) | 29% | 71% | |
| 3 (direct connection without filling the lumen) | 47% | 53% | |
| Total | 40% | 60% | |
| 2 (long, small diameter) | 1 (direct connection with filling the lumen) | 38% | 62% |
| 2 (three-way valve) | 19% | 81% | |
| 3 (direct connection without filling the lumen) | 39% | 61% | |
| Total | 33% | 67% | |
| 3 (short, large diameter) | 1 (direct connection with filling the lumen) | 35% | 65% |
| 2 (three-way valve) | 34% | 66% | |
| 3 (direct connection without filling the lumen) | 51% | 49% | |
| Total | 41% | 59% | |
| 4 (long, large diameter) | 1 (direct connection with filling the lumen) | 56% | 44% |
| 2 (three-way valve) | 22% | 78% | |
| 3 (direct connection without filling the lumen) | 63% | 37% | |
| Total | 48% | 52% | |
Air in the balloons for the different pressures.
| Air | |||
|---|---|---|---|
| Pressure | Method | Yes | No |
| Nominal pressure | 1 (direct connection with filling the lumen) | 38% | 62% |
| 2 (three-way valve) | 28% | 72% | |
| 3 (direct connection without filling the lumen) | 44% | 56% | |
| Total | 37% | 63% | |
| Rated burst pressure-1 | 1 (direct connection with filling the lumen) | 50% | 50% |
| 2 (three-way valve) | 25% | 75% | |
| 3 (direct connection without filling the lumen) | 58% | 42% | |
| Total | 45% | 55% | |
Total volume of air depending on the method.
| Volume Category | ||||
|---|---|---|---|---|
| High | Medium | Low | ||
|
| 1 (direct connection with filling the lumen) | 27% | 38% | 35% |
| 2 (three-way valve) | 32% | 32% | 36% | |
| 3 (direct connection without filling the lumen) | 39% | 28% | 33% | |
|
| 34% | 32% | 34% | |
Mean total volume depending on the method.
| Mean Total Volume (mm3) | ||
|---|---|---|
|
| 1 (direct connection with filling the lumen) | 1.12 |
| 2 (three-way valve) | 1.25 | |
| 3 (direct connection without filling the lumen) | 2.25 | |
|
| 1.64 | |