CONTEXT: A downward trend is being observed in the incidence of human immunodeficiency virus (HIV) infection in India due to strategic activities of National AIDS Control Organisation (NACO) in the last 24 years. Opt-out testing has consistently shown high seroprevalence in our tertiary care center. AIM: This study aims to audit opt-out testing and compare various commercial test kits used to detect HIV seroprevalence in patients in our tertiary care institute and suggest new algorithm for HIV testing in tertiary hospitals in India.Entities:
Keywords: Human immunodeficiency virus; molecular test; screening
Year: 2019 PMID: 31922102 PMCID: PMC6896380 DOI: 10.4103/ijstd.IJSTD_55_17
Source DB: PubMed Journal: Indian J Sex Transm Dis AIDS ISSN: 2589-0557
Various test kits and the rate of human immunodeficiency virus detection using these kits
| Kit | Principle | Detects | Solid phase | Sensitivity and specificity (%) | Reactive test in present study (%) |
|---|---|---|---|---|---|
| Microlisa | Fourth-generation EIA based on “Sandwich ELISA” | Antibodies to HIV-1 and/or HIV-2 and HIV-1 p24 Ag | The microtiter plate is coated with HIV envelope proteins gp41, C terminus of gp120 for HIV-1, and gp36 for HIV-2 and anti-HIV-1 p24 antibodies | 100 and 100 | 692/22,998 (3) |
| Immunocomb | Indirect solid-Phase EIA | Antibodies to HIV-1 and/or HIV-2 | Card with 12 projections (teeth). Each tooth is sensitized at 3 spots. The upper spot contains goat antibodies to human immunoglobulin (internal control); the middle and the lower spots contain HIV-2 and HIV-1 synthetic peptides | 100 and 99.4 | 159/6500 (2.4) |
| Genedia | Sandwich ELISA | Antibodies to HIV-1 gp41, HIV-1 group O gp41 and HIV-2 gp36 and, HIV-1 p24 antigen | Wells coated with recombinant HIV-1 gp41 Ag, recombinant HIV-1 group O gp41 Ag, recombinant HIV-2 gp36 Ag and monoclonal HIV-1 p24 antibodies | 100 and 99.7 | 6/190 (3.2) |
| Tridot | Rapid, visual immunoassay test | Antibodies (IgG) to HIV-1 and HIV-2 | HIV-1 and 2 antigens (gp41, C terminal of gp120 of HIV-1 and gp36 of HIV-2) immobilized on an immunofiltration membrane | 100 and 100 | 3/122 (2.5) |
| SD Bioline | Rapid immunochromatographic third generation of one step anti-HIV 1/2 test | Antibodies of all isotypes (IgG, M, A) specific to HIV-1 and HIV-2 | Membrane strip precoated with recombinant HIV-1 capture antigen (gp41 and p24) on test band-1 region and with recombinant HIV-2 capture antigen (gp36) on test band-2 region | 100 and 99.3 | 2/211 (0.95) |
HIV=Human immunodeficiency virus; EIA=Enzyme immunoassay; ELISA=Enzyme-linked immunosorbent assay
Figure 1Sequence of appearance of laboratory markers for human immunodeficiency virus-1 infection (Published in Laboratory Testing for the Diagnosis of human immunodeficiency virus infection: Updated Recommendations. Centre for disease control. June 27, 2014 http://stacks.cdc.gov/view/cdc/23447)
Figure 2Algorithm suggested by CDC for human immunodeficiency virus testing. (+): Reactive test result, (-): Nonreactive test result, NAT: Nucleic acid test
Figure 3Proposed testing algorithm for tertiary care centers in North India