| Literature DB >> 31909244 |
Evan Calkins1, Edvinas Pocius1, Gail Marracci1,2, Priya Chaudhary1.
Abstract
BACKGROUND: Fast, effective, and rapid processing of central nervous system (CNS) tissue with good preservation of myelin, especially in tissue from diseased mice, is important to many laboratories studying neurosciences. NEWEntities:
Keywords: Cell biology; Electron microscopy; Microwave; Neurology; Neuroscience; Ophthalmology; Optic nerve; Pathology; Pathophysiology; Plastic embedding; Spinal cord
Year: 2019 PMID: 31909244 PMCID: PMC6938874 DOI: 10.1016/j.heliyon.2019.e03036
Source DB: PubMed Journal: Heliyon ISSN: 2405-8440
Amounts of reagents needed for embedding.
| 1 specimen holder | 2 specimen holders | |
|---|---|---|
| ERL 4221 | 20 | 30 |
| DER 736 | 8 | 12 |
| NSA | 52 | 78 |
| DMAE | 0.80 | 1.20 |
| TOTAL | 80.80 | 121.20 |
| Mix with wooden spatula | ||
| EPON 812 | 28 | 42 |
| DDSA | 13 | 19.5 |
| NMA | 13.20 | 19.80 |
| DMP-30 | 2.14 | 3.21 |
| TOTAL | 56.34 | 84.51 |
| Mix with wooden spatula | ||
| TOTAL SPURR + EPON | 137.14 | 205.71 |
ERL 4221 is a cycloaliphatic expoxide resin; DER 736; NSA Nonenyl Succinic Anhydride; DMAE 2-Dimethylaminoethanol.
Epon 812; DDSA Dodecenyl Succinic Anhydride; NMA Methyl-5-Norbornene-2,3-Dicarboxylic Anhydride; DMP-30 2,4,6-Tris(dimethylaminomethyl)phenol.
Embedding protocol for the microwave.
| Steps | Description | user prompt | time (hr:min:sec) | Power (W) | Temp (oC) | Load Cooler | Vacuum |
|---|---|---|---|---|---|---|---|
| 1* | Cacodylate buffer rinse | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 2 | Osmium fixation | ON | 0:03:00 | 100 | 60 | AUTO | CYCLE |
| 3 | Osmium fixation | OFF | 0:02:00 | 0 | 60 | AUTO | CYCLE |
| 4 | Osmium fixation | OFF | 0:03:00 | 100 | 60 | AUTO | CYCLE |
| 5 | Osmium fixation | OFF | 0:02:00 | 0 | 60 | AUTO | CYCLE |
| 6 | Osmium fixation | OFF | 0:03:00 | 100 | 60 | AUTO | CYCLE |
| 7* | Distilled water rinse | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 8 | Uranyl acetate solution | ON | 0:02:00 | 100 | 60 | AUTO | CYCLE |
| 9 | Uranyl acetate solution | OFF | 0:02:00 | 0 | 60 | AUTO | CYCLE |
| 10 | Uranyl acetate solution | OFF | 0:02:00 | 100 | 60 | AUTO | CYCLE |
| 11 | 20% acetone | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 12 | 35% acetone | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 13 | 50% acetone | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 14 | 75% acetone | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 15 | 95% acetone | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 16* | 100% acetone | ON | 0:00:40 | 150 | 60 | AUTO | OFF |
| 17 | 3:1 acetone/resin | ON | 0:03:00 | 150 | 60 | AUTO | CYCLE |
| 18 | 2:1 acetone/resin | ON | 0:03:00 | 150 | 60 | AUTO | CYCLE |
| 19 | 1:1 acetone/resin | ON | 0:03:00 | 150 | 60 | AUTO | CYCLE |
| 20# | 100% Resin mixture -infiltration | ON | 0:03:00 | 150 | 60 | AUTO | CYCLE |
Please note: Steps 1,7,16* run 3 times; step 20# repeat 5–8 times as this may vary according to tissue size; steps 2–6 and 8–10 can be done in one dish thus no transferring in between steps. All steps with vacuum cycle will have the vacuum chamber shut tightly with the lid. The final resin polymerization step is done at 60–63 °C in a regular vacuum oven (VWR).
Fixation protocol for the microwave.
| Step | Time (min) | Power | Temp (oC) | Vacuum |
|---|---|---|---|---|
| 1 | 1 | 150 | 50 | Auto |
| 2 | 1 | 0 | 50 | Auto |
| 3 | 1 | 150 | 50 | Auto |
Repeat steps 1–3 twice.
Figure 1Microwave Pro + arrangement near the chemical hood. Notice that the exhaust in vented into the chemical hood.
Figure 2Spinal cord sections in 12 well Prep-Eze specimen holders a. before beginning embedding process; b. at the end of embedding process.
Figure 3Polymerized plastic blocks. a. optic nerves embedded in flat embedding mold; b. embedded spinal cord sections in embedding capsule.
Figure 4Light microscope images of spinal cord and optic nerve sections. Zeiss microscope with 63X objective was used for imaging. a. spinal cord b. optic nerve. Scale bar 5μm.
Figure 5EM images of spinal cord and optic nerve sections. The myelin is well preserved after using the fast and efficient microwave plastic embedding process. a. spinal cord section at 2900X, scale bar 2 μm; b. optic nerve section at 4800X, scale bar 1μm.