| Literature DB >> 31906315 |
Rama S Sista1, Rainer Ng1, Miriam Nuffer1, Michael Basmajian1, Jacob Coyne1, Jennifer Elderbroom1, Daniel Hull1, Kathryn Kay1, Maithri Krishnamurthy1, Christopher Roberts1, Daniel Wu1, Adam D Kennedy1, Rajendra Singh1, Vijay Srinivasan1, Vamsee K Pamula1.
Abstract
"Children are not tiny adults" is an adage commonly used in pediatrics to emphasize the fact that children often have different physiological responses to sickness and trauma compared to adults. However, despite widespread acceptance of this concept, diagnostic blood testing is an excellent example of clinical care that is not yet customized to the needs of children, especially newborns. Cumulative blood loss resulting from clinical testing does not typically impact critically ill adult patients, but can quickly escalate in children, leading to iatrogenic anemia and related comorbidities. Moreover, the tests prioritized for rapid, near-patient testing in adults are not always the most clinically relevant tests for children or newborns. This report describes the development of a digital microfluidic testing platform and associated clinical assays purposely curated to address current shortcomings in pediatric laboratory testing by using microliter volumes (<50 µL) of samples. The automated platform consists of a small instrument and single-use cartridges, which contain all reagents necessary to prepare the sample and perform the assay. Electrowetting technology is used to precisely manipulate nanoliter-sized droplets of samples and reagents inside the cartridge. To date, we have automated three disparate types of assays (biochemical assays, immunoassays, and molecular assays) on the platform and have developed over two dozen unique tests, each with important clinical application to newborns and pediatric patients. Cell lysis, plasma preparation, magnetic bead washing, thermocycling, incubation, and many other essential functions were all performed on the cartridge without any user intervention. The resulting assays demonstrate performance comparable to standard clinical laboratory assays and are economical due to the reduced hands-on effort required for each assay and lower overall reagent consumption. These capabilities allow a wide range of assays to be run simultaneously on the same cartridge using significantly reduced sample volumes with results in minutes.Entities:
Keywords: albumin; bilirubin; creatinine; cytomegalovirus; digital microfluidics; electrowetting; glucose-6-phosphate dehydrogenase; insulin; laboratory; nanoliter; neonatal intensive care unit (NICU); pediatrics; thyroid
Year: 2020 PMID: 31906315 PMCID: PMC7169462 DOI: 10.3390/diagnostics10010021
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Requirements for routine coagulation testing.
| Analyte | Reference Method [ | Expected Turnaround Time | Minimum Volume of Plasma/Serum | Estimated Volume of Whole Blood | Additional Pre-Shipping Sample Preparation |
|---|---|---|---|---|---|
| Heparin Anti-Xa | Labcorp 117101 (activity assay) | 1–3 days | 1.0 mL | 2.0 mL | Centrifugation right after collection and frozen shipping |
| Factor VIII activity | Labcorp 500192 (activity assay) | 3–5 days | 0.5 mL | 1.0 mL | Same as above |
| ATIII activity | Labcorp 015040 (activity assay) | 2–3 days | 1.0 mL | 2.0 mL | Same as above |
| Von Willebrand factor antigen | Labcorp 086280 (immunoassay) | 1–3 days | 1.0 mL | 2.0 mL | Same as above |
| Protein S antigen | Labcorp 164517 (immunoassay) | 2–3 days | 2.0 mL | 4.0 mL | Same as above |
| Factor V Leiden Mutation Analysis | Labcorp 511154 (nucleic acid assay) | 5–7 days | n/a | 3.0 mL | None |
| Factor II (Prothrombin), DNA Analysis | Labcorp 511162 (nucleic acid assay) | 5–7 days | n/a | 3.0 mL | None |
Figure 1Rendering of the main components of the digital microfluidic instrument.
Figure 2The single-use cartridge is comprised of a printed circuit board and a plastic top plate. Notable features include the sealed liquid reagent module, sample loading port, and dried reagents (not shown).
Figure 3(A) Composite photograph of plasma separation on cartridge. (B) Hemoglobin levels in the plasma obtained from biochemical or centrifugation separation methods are not statistically different as measured by paired t-test. This study demonstrates that biochemical plasma separation does not induce excessive stress on the red blood cells (RBCs). (C) Whole blood from a single donor was lysed using either standard osmotic methods on the bench (left) or via osmotic lysis on the digital microfluidic (DMF) cartridge (right). The scatter at 775 nm is comparable in both samples, indicating that the DMF lysis method is as effective as the standard bench osmotic shock technique. (D) Sample input volume (x-axis) versus impedance (y-axis); circle: fail, triangle: pass. AU: arbitrary units.
Summary of assays adapted to the diagnostic DMF platform.
| Sample Matrix | Immunoassays | Molecular | Chemical/Enzymatic Assays | |||
|---|---|---|---|---|---|---|
| Competitive | Sandwich | Genomic (Eukaryotic) | Infectious Disease (Prokaryotic) | Functional | Biochemical | |
| Blood | Free T4, cortisol | Insulin, human growth hormone, protein S, von Willebrand factor, thyroid stimulating hormone, creatine kinase MM | Factor II mutation, factor V Leiden | HIV | G6PD, antithrombin III, protein C, factor VIII, galactose-1-phosphate uridyltransferase | Total bilirubin, albumin, unbound bilirubin, glucose, beta-hydroxybutyrate, free fatty acids, factor Xa, phenylalanine, creatine kinase, glutamine, glutamate, ammonia |
| Urine | Fentanyl | NGAL, Cystatin C | Creatinine | |||
| Saliva | CMV | |||||
Figure 4Method comparison results of the DMF assays (y-axis) to comparator laboratory assays performed at LabCorp (x-axis) for total bilirubin (TBil) (A), albumin (B), and glucose-6-phosphate dehydrogenase (G6PD) (D), or using an in vitro diagnostic comparator method for unbound bilirubin (C).
Figure 5Analysis of urine samples using DMF. Method comparison using the DMF method (y-axis) for creatinine versus the comparator method using mass spectrometry (x-axis).
Figure 6Immunoassay results for insulin (A), free T4 (B) and neutrophil gelatinase-associated lipocalin (NGAL) (C) on the DMF cartridge. (A,B) are sandwich and competitive immunoassays, respectively, using plasma calibrant samples; (C) is a sandwich immunoassay using urine calibrant samples.
Figure 7Cytomegalovirus (CMV) polymerase chain reaction (PCR) assay using a prototype DMF cartridge. Successful amplification of CMV (10, 1500, and 225,000 copies/µL) is achieved within 5 min. FU: fluorescence units.