| Literature DB >> 31906242 |
Adrien Roumiguières1,2, Said Kinani1, Stéphane Bouchonnet2.
Abstract
Membrane-introduction mass spectrometry (MIMS) has been presented as one of the promising approaches for online and real-time analysis of monochloramine (NH2Cl) in diverse matrices such as air, human breath, and aqueous matrices. Selective pervaporation of NH2Cl through the introduction membrane overcomes the need for sample preparation steps. However, both the selectivity and sensitivity of MIMS can be affected by isobaric interferences, as reported by several researchers. High-resolution mass spectrometry helps to overcome those interferences. Recent miniaturization of Fourier transform-ion cyclotron resonance-mass spectrometry (FT-ICR MS) technology coupled to the membrane-introduction system provides a potent tool for in field analysis of monochloramine in environmental matrices. Monochloramine analysis by MIMS based FT-ICR MS system demonstrated decomposition into ammonia. To further clarify the origin of this decomposition, headspace analyses after bypassing the membrane were undertaken and showed that monochloramine decomposition was not exclusively related to interactions within the membrane. Adsorption inside the MIMS device, followed by surface-catalyzed decomposition, was suggested as a plausible additional mechanism of monochloramine decomposition to ammonia.Entities:
Keywords: Fourier transform-ion cyclotron resonance; adsorption; membrane-introduction mass spectrometry; monochloramine; surface-catalyzed decomposition
Year: 2019 PMID: 31906242 PMCID: PMC6982842 DOI: 10.3390/s20010247
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Diagram of the inlet system in the BTrap™ device.
Figure 2Influence of flow rate on Membrane Introduction Mass Spectrometry (MIMS) signal intensity. Ions with m/z values of 51.995 (NH335Cl+: Blue) and 53.992 (NH337Cl+: Red) were selected to monitor monochloramine. Experiments were conducted in duplicate.
Figure 3Mass spectrum obtained by MIMS for aqueous solution of monochloramine at a concentration of 70 mg L−1 (as Cl2) and pH between 7.0–7.2. The inversed spectrum at the bottom is from the ultrapure water (as blank).
Figure 4Kinetics of ion formation from a 20 mM NH4Cl solution (relative standard deviation (RSD) < 15%, n = 3). The two time-points, t0 and tend, represent the beginning of headspace introduction and time of solution removal, respectively.
Figure 5Stepwise chlorination of a 20 mM NH4Cl solution with different molar Cl:N ratios in headspace mode.
Figure 6Relative ion intensities (intensity of each ion with respect to the sum of all ions) measured from a 20 mM NH2Cl solution as a function of proton transfer reaction time in headspace mode. Standard deviations are shown as error bars (n = 3).
Figure 7Schematic representation of the adsorption/desorption and surface catalyzed decomposition phenomena.