Literature DB >> 31899806

Comprehensive radiation and imaging isocenter verification using NIPAM kV-CBCT dosimetry.

Kiran Pant1, Chibuike Umeh2,3, Mark Oldham4, Scott Floyd4, Will Giles4, Justus Adamson4.   

Abstract

PURPOSE: To develop and demonstrate a comprehensive method to directly measure radiation isocenter uncertainty and coincidence with the cone-beam computed tomography (kV-CBCT) imaging coordinate system that can be carried out within a typical quality assurance (QA) time slot.
METHODS: An N-isopropylacrylamide (NIPAM) three-dimensional (3D) dosimeter for which dose is observed as increased electron density in kV-CBCT is irradiated at eight couch/gantry combinations which enter the dosimeter at unique orientations. One to three CBCTs are immediately acquired, radiation profile is detected per beam, and displacement from imaging isocenter is quantified. We performed this test using a 5 mm diameter MLC field, and 7.5 and 4 mm diameter cones, delivering approximately 16 Gy per beam. CBCT settings were 1035-4050 mAs, 80-125 kVs, smooth filter, 1 mm slice thickness. The two-dimensional (2D) displacement of each beam from the imaging isocenter was measured within the planning system, and Matlab code developed in house was used to quantify relevant parameters based on the actual beam geometry. Detectability of the dose profile in the CBCT was quantified as the contrast-to-noise ratio (CNR) of the irradiated high-dose regions relative to the surrounding background signal. Our results were compared to results determined by the traditional Winston-Lutz test, film-based "star shots," and the vendor provided machine performance check (MPC). The ability to detect alignment errors was demonstrated by repeating the test after applying a 0.5 mm shift to the MLCs in the direction of leaf travel. In addition to radiation isocenter and coincidence with CBCT origin, the analysis also calculated the actual gantry and couch angles per beam.
RESULTS: Setup, MV irradiation, and CBCT readout were carried out within 38 min. After subtracting the background signal from the pre-CBCT, the CNR of the dosimeter signal from the irradiation with the MLCs (125 kVp, 1035 mAs, n = 3), 7.5 mm cone (125 kVp, 1035 mAs, n = 3), and 4 mm cone (80 kVp, 4050 mAs, n = 1) was 5.4, 5.9, and 2.9, respectively. The minimum radius that encompassed all beams calculated using the automated analysis was 0.38, 0.48, and 0.44 mm for the MLCs, 7.5 mm cone, and 4 mm cone, respectively. When determined manually, these values were slightly decreased at 0.28, 0.41, and 0.40 mm. For comparison, traditional Winston-Lutz test with MLCs and MPC measured the 3D isocenter radius to be 0.24 mm. Lastly, when a 0.5 mm shift to the MLCs was applied, the smallest radius that intersected all beams increased from 0.38 to 0.90 mm. The mean difference from expected value for gantry angle was 0.19 ± 0.29°, 0.17 ± 0.23°, and 0.12 ± 0.14° for the MLCs, 7.5 mm cone, and 4 mm cone, respectively. The mean difference from expected for couch angle was -0.07 ± 0.28°, -0.08 ± 0.66°, and 0.04 ± 0.25°.
CONCLUSIONS: This work demonstrated the feasibility of a comprehensive isocenter verification using a NIPAM dosimeter with sub-mm accuracy which incorporates evaluation of coincidence with imaging coordinate system, and may be applicable to all SRS cones as well as MLCs.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  3D dosimetry; isocenter; radiosurgery; winston-Lutz

Year:  2020        PMID: 31899806     DOI: 10.1002/mp.14008

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  A 3D star shot to determine the gantry, collimator, and couch axes positions.

Authors:  Robert Corns; Kaida Yang; Mason Ross; Shiva Bhandari; Makunda Aryal; Peter Ciaccio
Journal:  J Appl Clin Med Phys       Date:  2022-04-29       Impact factor: 2.243

2.  Fast Isocenter Determination Using 3D Polymer Gel Dosimetry with Kilovoltage Cone-Beam CT Reading and the PolyGeVero-CT Software Package for Linac Quality Assurance in Radiotherapy.

Authors:  Piotr Maras; Marek Kozicki
Journal:  Materials (Basel)       Date:  2022-09-30       Impact factor: 3.748

3.  Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT.

Authors:  Edward T Cullom; Yuqing Xia; Kai-Cheng Chuang; Zachary W Gude; Yana Zlateva; Justus D Adamson; William M Giles
Journal:  J Appl Clin Med Phys       Date:  2021-06-24       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.