Literature DB >> 31897601

High density mapping and catheter ablation of atrial tachycardias in adults with congenital heart disease.

Ulrich Krause1, Matthias J Müller2, Canan Stellmacher2, David Backhoff2, Heike Schneider2, Thomas Paul2.   

Abstract

AIMS: We used a new grid-style multi-electrode mapping catheter (Advisor™ HD Grid, Abbott) and investigated its use for high density mapping of atrial tachycardias in adult patients with congenital heart disease. PATIENTS AND METHODS: All patients with congenital heart disease who had mapping of atrial tachycardias using the new grid-style catheter between March 2018 and April 2019 were included.
RESULTS: A total of 24 adult patients had high density mapping of atrial tachycardias using the grid-style multi-electrode catheter. Mean procedure duration was 207 ± 72 min., mean fluoroscopy time was 7.1 ± 7.9 min. In patients with right atrial substrates, fluoroscopy time was shorter compared to biatrial or left atrial substrates (0.9 ± 2.2 min for right atrial substrates, n = 19 vs. 6.3 ± 8.3 min for left atrial substrates, n = 2 and 7.5 ± 4.3 min for biatrial substrates, n = 3, p = 0.01). A mean number of 14.814 ± 10.140 endocardial points were collected and 2.319 ± 1244 points were finally used to characterize the tachycardia. Procedural success was achieved in 21/24 (88%) subjects and partial success in 2/24 (8%) patients. Recurrence rate was low (12.5%). In one patient, radiofrequency ablation within the cavotricuspid isthmus resulted in occlusion of a branch of the right coronary artery. No complications related to the use of the mapping catheter itself occurred.
CONCLUSION: High density mapping of AT using the grid-style catheter showed promising results with respect to procedural and midterm outcome and fluoroscopy time. Using the grid-style catheter might offer advantages compared to other multi-electrode catheters used for high density mapping of AT in patients with CHD.

Entities:  

Keywords:  Atrial tachycardia; Catheter ablation; Congenital heart disease; High density mapping; Multi-electrode catheter

Mesh:

Year:  2020        PMID: 31897601     DOI: 10.1007/s00392-019-01592-x

Source DB:  PubMed          Journal:  Clin Res Cardiol        ISSN: 1861-0684            Impact factor:   5.460


  1 in total

1.  Koch's triangle voltage mapping for cryoablation of slow pathway in children: preliminary data of a novel high-density technique.

Authors:  Fabrizio Drago; Pietro Paolo Tamborrino; Luigina Porco; Marta Campisi; Veronica Fanti; Roberta Annibali; Massimo Stefano Silvetti
Journal:  J Interv Card Electrophysiol       Date:  2021-10-25       Impact factor: 1.900