| Literature DB >> 31896918 |
Ravi C Dara1, Aseem Kumar Tiwari2, Subhasis Mitra2, Deviprasad Acharya2, Geet Aggarwal2, Dinesh Arora2, Gunjan Bhardwaj2.
Abstract
INTRODUCTION: Semi-automated equipment using Column agglutination technology (CAT) is widely used where centrifugation and incubation are automated but substantial amount of the work is still executed manually. Larger laboratories are moving towards automation to eliminate errors, reducing exposure to bio-hazardous samples, assuring traceability, reliability, turnaround time (TAT) and throughput. Moving towards automation and greater reliability, we therefore, decided to install an automated immunohematological (IH) analyzer "Vision". In this study we evaluated reliability and performance before clearing "Vision" for routine use.Entities:
Keywords: Column agglutination technology; immunohematology; pretransfusion testing
Year: 2019 PMID: 31896918 PMCID: PMC6910038 DOI: 10.4103/ajts.AJTS_116_17
Source DB: PubMed Journal: Asian J Transfus Sci ISSN: 0973-6247
Analysis of blood group discrepancy samples using Vision and BioVue system
| Discrepancy type | Known blood group discrepancy | Results (vision IH analyzer) | Results (BioVue IH system) |
|---|---|---|---|
| Type 1 | Forward O/reverse | Forward | Forward |
| Forward A/reverse | Forward | Forward | |
| Type 2 | Weak A positive (subgroup of A) | A Rh D positive | O Rh D positive |
| Weak A positive (subgroup of A) | O Rh D positive | O Rh D positive | |
| Type 4 | Auto-control positive | Auto-control positive | Auto-control positive |
| Alloantibody positive | Alloantibody positive | Alloantibody positive | |
| A2 with anti-A1 | A2 with anti-A1 | A2 with anti-A1 |
Antibody identification results using Vision and BioVue system
| Known antibody | n | Vision IH analyzer | BioVue system | ||
|---|---|---|---|---|---|
| Correctly identified | Not/incorrectly identified | Correctly identified | Not/incorrectly identified | ||
| Anti D | 13 | 13 | 0 | 13 | 0 |
| Weak anti D | 11 | 11 | 0 | 11 | 0 |
| Anti Fya | 1 | 1 | 0 | 1 | 0 |
| Anti Jka | 1 | 1 | 0 | 1 | 0 |
| Anti M | 5 | 5 | 0 | 5 | 0 |
| Weak anti M | 1 | 1 | 0 | 0 | 1 |
| Anti E + S | 1 | 1 | 0 | 1 | 0 |
| Anti C + K | 1 | 1 | 0 | 1 | 0 |
| Anti S | 1 | 1 | 0 | 1 | 0 |
| Anti K | 1 | 1 | 0 | 1 | 0 |
| Anti C | 1 | 1 | 0 | 1 | 0 |
| Anti c | 1 | 1 | 0 | 1 | 0 |
| Anti E | 1 | 1 | 0 | 1 | 0 |
| Anti e | 1 | 1 | 0 | 1 | 0 |
| Anti Jkb | 1 | 1 | 0 | 1 | 0 |
| Anti s | 1 | 1 | 0 | 1 | 0 |
| Anti P | 1 | 1 | 0 | 1 | 0 |
| Anti D + C | 1 | 1 | 0 | 1 | 0 |
| Anti JKb + C | 1 | 1 | 0 | 1 | 0 |
Figure 1Comparison of Turn Around Time (TAT) between Vision and Reference system for different test profiles
Figure 2Stress test