| Literature DB >> 31889773 |
Seung-Yong Hahn1, Juan Bascuñán1, Woo-Seok Kim1, Emanuel S Bobrov1, Haigun Lee2, Yukikazu Iwasa1.
Abstract
This paper describes two general methods: field map processing and NMR lineshape analysis, that we are currently using for the on-going research of field improvement techniques specifically applied to our 700 MHz LTS/HTS NMR magnet. The validity of the methods is verified with comparison between calculations and experiments. Also, a simple but effective analysis has been used to identify the principal source of a remanent magnetic field measured in the bore of the 700 MHz LTS/HTS NMR magnet: the Screening Current induced Field (SCF) from the 100-MHz HTS insert, comprised of 48 double pancake coils, each wound with Bi2223 tape.Entities:
Keywords: Field mapping; LTS/HTS NMR magnet; NMR lineshape; screening currents induced field (SCF)
Year: 2008 PMID: 31889773 PMCID: PMC6936755 DOI: 10.1109/TASC.2008.921225
Source DB: PubMed Journal: IEEE Trans Appl Supercond