How Center of Rotation Changes and What Affects These After Cervical Arthroplasty: A Systematic Review and Meta-analysis.

Hongpeng Sang1, Wei Cui1, Dacheng Sang1, Ze Guo1, Baoge Liu2.   

Abstract

PURPOSE: To determine how center of rotation (COR) changes and what affects changes in COR after cervical arthroplasty.
METHODS: A systematic literature review of in vivo clinical studies comparing the location of the COR before and after cervical arthroplasty with different artificial prostheses was performed. Meta-analysis was performed using a fixed effects model where appropriate.
RESULTS: A systematic review of the PubMed, EMBASE, and Cochrane Library databases was conducted. We initially identified 267 studies, of which 14 involved in vivo kinematics studies evaluating COR following cervical arthroplasty. We found that at the last follow-up, the COR location shifted anteriorly in patients from 4 studies including 85 segments, superiorly in patients from 4 studies including 98 segments, anterior-superiorly in patients from 4 studies including 290 segments, and anterior-inferiorly in patients from 1 study including 272 segments after cervical arthroplasty. The COR location showed no significant change in patients from 5 studies including 106 segments after cervical arthroplasty. Changes in COR showed a certain trend after cervical arthroplasty with different types of prostheses.
CONCLUSIONS: Prosthesis design affects changes in COR after cervical arthroplasty. If a constrained or semiconstrained prosthesis is chosen (2-piece implant, ball-and-socket, or ball-in-trough design), the COR location tends to shift anteriorly and/or superiorly, whereas if a nonconstrained prosthesis is chosen (3-piece implant, mobile nucleus design), the COR tends to keep the same location as preoperation. In addition, the position of the prosthesis in the intervertebral space also can affect changes in COR after cervical arthroplasty.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Center of rotation; Cervical arthroplasty; Clinical study; In vivo

Mesh:

Year:  2019        PMID: 31887466     DOI: 10.1016/j.wneu.2019.12.111

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  2 in total

1.  Radiological Changes in Adjacent and Index Levels after Cervical Disc Arthroplasty.

Authors:  Jun Jae Shin; Kwang-Ryeol Kim; Dong Wuk Son; Dong Ah Shin; Seong Yi; Keung-Nyun Kim; Do-Heum Yoon; Yoon Ha
Journal:  Yonsei Med J       Date:  2022-01       Impact factor: 2.759

2.  Mid-long-term follow-up of operated level kinematics after single-level artificial cervical disc replacement with Bryan disc.

Authors:  Chuanhong Li; Xing Yu; Yang Xiong; Yongdong Yang; Fengxian Wang; He Zhao
Journal:  J Orthop Surg Res       Date:  2022-03-09       Impact factor: 2.359

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.