Literature DB >> 31873651

Improved spherical mirror multipass-cell-based interband cascade laser spectrometer for detecting ambient formaldehyde at parts per trillion by volume levels.

Bo Fang, Nana Yang, Weixiong Zhao, Chunhui Wang, Weijun Zhang, Wei Song, Dean S Venables, Weidong Chen.   

Abstract

We report the development of an improved spherical mirror multipass-cell-based interband cascade laser (ICL) spectrometer for ambient formaldehyde (HCHO) detection. The multipass cell consists of two easily manufactured spherical mirrors that are low cost, and have a simple structure, large mirror area utilization, and dense spot pattern. Optical interference caused by the multipath cell was largely reduced, resulting in good sensitivity. Using wavelength modulation spectroscopy (WMS), a detection precision (${1} \sigma $1σ) of 51 pptv in 10 s was achieved with an absorption pathlength of 96 m, which compared favorably with the performance of other state-of-the-art instruments. The precision can be further improved by using a long absorption pathlength configuration and by removing fringe-like optical noise caused by the collimation lens. Ambient application of the developed spectrometer was demonstrated.

Entities:  

Year:  2019        PMID: 31873651     DOI: 10.1364/AO.58.008743

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  Towards Interband Cascade lasers on InP Substrate.

Authors:  Krzysztof Ryczko; Janusz Andrzejewski; Grzegorz Sęk
Journal:  Materials (Basel)       Date:  2021-12-22       Impact factor: 3.623

  1 in total