Literature DB >> 31866552

Assessment of diagnostic accuracy of SanketLife - A wireless, pocket-sized ECG biosensor, in comparison to standard 12 lead ECG in the detection of cardiovascular diseases in a tertiary care setting.

Siva Kumar1, C M Nagesh1, Manmohan Singh2, Anbu Pandian3, David Delurgio4, Bobby Khan5, Robin Chaudhary6, Prashant Gupta7.   

Abstract

BACKGROUND: The SanketLife is a low cost, portable, pocket sized 12 lead ECG mechanised by SanketLife app running on compatible iOS and Android phones that connect wirelessly via Bluetooth technology to the device.
OBJECTIVE: The current study was conducted to assess the diagnostic accuracy of SanketLife ECG in comparison to standard 12 lead ECG (GE-2000) in detection of cardiovascular diseases. RESEARCH DESIGN AND METHODS: This was a prospective diagnostic test accuracy trial conducted in outpatient settings of a tertiary cardiac care centre in India. A total of 100 patients, attended cardiology OPD, were included in the study. Consecutive ECGs were taken by 12 lead standard ECG as well as by SanketLife ECG. Diagnostic accuracy variables such as sensitivity, specificity, negative and positive predictive value, negative and positive likelihood ratios were estimated. Ethical permission was taken from the Institutional ethical committee. RESULTS &
CONCLUSION: The analysis showed a high degree of agreement and accuracy of SanketLife in detecting major cardiovascular conditions (Major Minnesota codes) such as Left and right bundle branch block, ST-segment elevation and ST-segment depression, AV conduction block. SanketLife showed high sensitivity (98.15%) and specificity (100%) in diagnosing major cardiovascular conditions.
Copyright © 2019 Indian Heart Rhythm Society. All rights reserved.

Entities:  

Keywords:  12-Lead; Accuracy; Agreement; Minnesota; SanketLife; Sensitivity

Year:  2019        PMID: 31866552      PMCID: PMC7082670          DOI: 10.1016/j.ipej.2019.12.011

Source DB:  PubMed          Journal:  Indian Pacing Electrophysiol J        ISSN: 0972-6292


Background

Cardiovascular diseases in India

The age-standardized cardiovascular disease mortality rate in India is higher than the global average (272 vs 235 per 100,000 population) [1]. It is not only the high incidence rate of cardiovascular diseases in the Indian population, but they also get affected at a younger age. Cardiovascular diseases manifest in Indian population on an average a decade earlier than western countries and impact working age group substantially in terms of deaths and morbidity [2,3]. The magnitude of the financial burden of Cardiovascular diseases in India is also enormous. According to a health economics study, hospitalizations with CVD resulted in 12% higher odds of incurring catastrophic spending and 37% greater odds of falling into poverty in India in comparison to communicable diseases [4].

Delay in care and field triage

High incidence rate along with poor outcome indicates failure of the preventive measure at all levels from primary to tertiary in preventing cardiovascular mortality and morbidity. It is not only the sedentary lifestyle, inappropriate diet, the high prevalence of diabetes and hypertension but also poor management of acute cardiac events such as myocardial infarction, which set back efforts towards reducing premature mortality associated with CVDs. According to a multi-centre study [5] conducted in Punjab (India), 42% of all patients presenting with STEMI (ST-elevation myocardial infarction) at five different tertiary cardiac care centres were late by more than 6 h because of the pre-hospital delay. Delay was significantly associated with factors such as being elderly, rural residence and illiteracy. Recognizing symptoms as cardiac in origin, hospital as first medical contact, pre-hospital ECG was associated with shorter delays. Another study [6] by Ong Me advocated out of hospital ECG and transmission to higher centres as the best practice for chest pain management. This study estimated a significant reduction in the door to balloon times in patients with acute STEMI with the help of out of hospital 12 lead ECG recording and transmission. A tertiary care hospital-based study [7] in Srinagar estimated that 44.7% of acute STEMI reported late to the hospital because of misdiagnosis by local healthcare providers. Rural residence, referral, and transport delay were among other important reasons. S. Mehta et al. in his article [8] discussed seven system barriers in the management of acute STEMI in low- and middle-income countries especially in the context of India. Unavailability of lowcost ECG machines, lack of training of general physicians in its interpretation, risk stratification/triage at the primary level, the transmission of vitals and ECG findings to the corresponding tertiary care centres were highlighted with other reasons such as lack of insurance. Use of telemedicine, cheap ECG, empowering general physicians by training and continuous medical education (CME) should be the part of the broader strategy to improve the overall outcome of patients. A metanalysis [9] in its findings concluded that pre-hospital triage and telemedicine are associated with a nearly halved time to treatment in STEMI. Other studies in India [10] and outside [11] has shown significant improvement in the management of patients with MI by a well-coordinated healthcare system, where timely triage at periphery results in better outcomes.

SanketLife role as a biosensor in early detection of myocardial infarction

Biosensors along with artificial intelligence and machine learning have been projected for long as the future of cardiac care [12]. Marjan Gusev in his article [13] presented the framework of using ECG as a biosensor in as heart attack alerting system. This system consisted of a small portable ECG biosensor placed at the patient’s body and an application installed in the smartphone, which can receive ECG signals and transmit them to the data centre and also a web application on the cloud. The data centre can detect any abnormal heart activity, while the cloud-based application can convey it to the doctors in charge of providing 24-h medical services. Another similar article [14] also described the detection of heart attack by using a wearable ECG sensor and a mobile phone. SanketLife ECG works on similar line as a biosensor in the detection of electrical abnormalities and communicating those to the data centre through an application installed in the smartphone.

Rationale of the study

According to Guyatt and colleagues, “Diagnostic technology should be disseminated only if they are less expensive, produce fewer untoward effects, at least as accurate as existing technologies, eliminate the need for other diagnostic interventions, without loss of accuracy or lead to the institution of effective therapy”. It is evident that portable, low cost 12 lead ECG can be very helpful in the triage of patients with chest pain in peripheral areas. In countries like India, where more than half of the population lives in villages, far from the tertiary care centres, cheap ECG along with telemedicine will be a boon for millions of people. The current study was done to assess the diagnostic accuracy SanketLife device against standard 12 lead ECG.

Aim and objective

Primary Objective of the study was to establish diagnostic accuracy of SanketLife (A portable, wireless 12 lead ECG) against standard 12 lead ECG in the detection of cardiovascular diseases in a tertiary outpatient care setting.

Research design and methods

This was a prospective diagnostic accuracy trial, where participants underwent for both type of ECG (a standard 12 lead ECG as well as SanketLife) consecutively. The primary objective of the study was to ascertain the diagnostic accuracy and reliability of the SanketLife ECG in comparison to standard 12 lead ECG. Randomization of subjects was not done. The study was conducted in a tertiary care setting. A total of 100 patients was included in the study at a tertiary care setting. Though [15] it was convenient sampling but according to a study nomogram calculates a sample size of 91 at sensitivity of 0.97, prevalence of 0.25 and absolute precision of 0.07. A sample size of 100 has Patients included in the study either came to the cardiac tertiary care centre with symptoms suggestive of cardiovascular disease or have a history of it. All patients who were referred for 12 lead standard ECG, also got their consecutive ECG done with SanketLife. (Fig. 1) ECG reports were stored at the company’s cloud. ECG data were retrieved, once the sample size was reached. All reports were coded according to the Minnesota coding protocol.
Fig. 1

Study design.

Study design. ECG variables namely Heart Rate, PR, QRSD, QT & QTc were monitored, displayed and saved in our system and compared against certain heart abnormalities visual against traditional 12 LEAD ECG machine.

Quality control and quality assurance procedures

The clinical trial is conducted (Data management, documentation and reporting) in compliance with the protocol, ICH GCP guidelines and in line with other regulatory protocols. The ICH GCP guideline was developed with the consideration of the current good clinical practices of the European Union and countries like Japan, and the United States, Australia, Canada, the Nordic countries and the World Health Organization (WHO).

SanketLife ECG device

SanketLife is an ECG monitor which is pocket-able, leadless and can support up to 12 lead. The proposed device is mechanised by a SanketLife app running on compatible iOS and Android phones that connect wirelessly via Bluetooth technology to the device. The 12 lead ECG can be measured anytime, anywhere and the reports can be easily downloaded and instantly shared with the healthcare professional for immediate clinical advice. The ECG reports are also saved in the cloud for future reference. The device mainly consists of electronics such as sensors, comparators, filters and, Bluetooth chip.

Statistical analysis

Following end, outcomes were calculated for SanketLife in comparison to standard 12 lead ECG (GE 2000) for all Minnesota codes (disease conditions) separately. Specificity and sensitivity Positive and negative predictive value Positive and the negative likelihood ratio Disease prevalence

Inclusion criteria

Male or female subjects aged 18 years or more, willing and able to give informed consent. Diagnosed with required disease/severity/symptoms, any specific assessment criteria for these, or be in good health Participants have clinically acceptable laboratory and ECG done at the same time and place along with SanketLife ECG

Exclusion criteria

Significant renal or hepatic impairment and scheduled elective surgery or other procedures Terminally ill and any other significant disease or disorder which, in the opinion of the Investigator, may either put the participants at risk because of participation in the study, or may influence the result of the study, or the participant’s ability to participate in the study.

Ethical considerations and participant’s confidentiality

Ethical permission was taken from the institutional ethics committee of Shri Jayadeva Institute of Cardiology, Bangalore. Confidentiality of participants was maintained by de-identifying personal identifiers and only anonymized data were used in the analysis. All data and related information were stored securely and only accessible to trial staff and principal investigator. Written informed consent was taken from all patients before participating in the study.

Results: data analysis

The total sample size on which analysis conducted, was 91 because of the poor quality of reports. All subjects were male with a mean age of 52.02 years (SD: 14.04) and median of 55. Age of subjects ranges from 24 to 91 years. Matching was done between Minnesota codes assigned to ECG findings of SanketLife and standard 12 lead ECG. Individual codes (a few cases had more than one code) were matched. Abnormal Q waves (Minnesota code:1-1-6) were infrequent and reported in only one case by both types of ECGs. (Table 1) Left axis deviation was reported in two cases with 100% matching of results. In Minnesota code 3 (High amplitude R waves) similar findings were reported by SanketLife as of standard 12 lead ECG in 9 cases. Though in 2 cases, where 12 lead ECG shows normal R waves, SanketLife showed high amplitude R waves.
Table 1

ECG (Standard and SanketLife) findings (Minnesota code wise).

ECG findingsNumberOf subjectsMinnesota coding (SanketLife)Minnesota coding (12 lead standard)MatchingComme nt
Q waves (Code:1)
Small Initial R wave in V1 and V2 lead11-1-61-1-6Yes
Axis deviation (2)
Left22–12–1Yes
High Amplitude R waves (3)
Left type83–13–1Yes
Left type23–1NormalNo
Right type13–23–2Yes
ST depression (4)
STJ depression ≥ 2.0 mm and ST segment horizontal or downward sloping in any of leads I, aVL, or V684-1-14-1-1Yes
STJ depression ≥ 1.0 mm but <2.0 mm, and ST segment horizontal or downward sloping in any of leads I, aVL, or V6.24-1-24-1-2Yes
STJ depression ≥1.0 mm but <2.0 mm, and ST segment horizontal or downward sloping in any of leads I, aVL, or V6104–44–4Yes
STJ depression ≥1.0 mm and ST segment upward sloping or Ushaped, in any of leads I, aVL, or V6.24–4No upward sloping ST segmentNo
T wave negativity (5)
Negative T-wave in Lead I, II &V5 V6275–25–2Yes
Flat T wave in V1 and Positive T wave in V225–35–3Yes
Flat T wave in V1 and Positive T wave in V225–31.Negative T wave in V1 & Flat T wave in V2 2. Normal R wave3No
1No 5-35–3No
AV conduction defect (6)
Wenckebach’s Phenomenon (P-R interval increasing from beat to beat until QRS and T dropped)36-2-36-2-3Yes
Ventricular blocks (7)
Left bundle branch block27-1-17-1-1Yes
Right bundle branch block27-2-17-2-1Yes
Intraventricular conduction delay27–47–4Yes
Arrythmias (8)
Atrial or junctional premature beats18-1-1NormalNo
Atrial or junctional premature beats1Normal8-1-1No
Ventricular premature beats38-1-28-1-2Yes
Ventricular premature beats58-1-2NormalNo
Atrial fibrillation (Persistent)18-3-18-3-1Yes
Low QRS amplitude (9-1)19–19–1Yes
ST segment elevation (9-2)
169–29–2Yes
Hyperacute T wave (9-5)
79–59–5Yes
Normal ECGs
10NormalNormalYes
ECG (Standard and SanketLife) findings (Minnesota code wise). ST depression (4-1-1, 4-1-2, 4-4) was reported in 22 ECGs by the standard as well as SanketLife with mismatching in two ECGs where 12 lead standard ECG was normal while SanketLife found ST junction depression (Code 4-4). Out of 32 cases, the pattern of T waves were similar in 29 cases in both type of ECGs. Atrio-ventricular defects and ventricular block were reported in 9 cases with 100% matching. ST-segment elevation and Hyperacute T waves also reported by both type of ECGs among 23 cases with 100% matching. In Minnesota coding category 8 (Arrhythmia), there is a total of 11 cases. There was a lack of agreement in both the cases of atrial premature beats and four cases of premature ventricular beats.

Major and minor ECG abnormalities (Table 2)

Minnesota coding has been extensively used in population studies of ECG abnormalities and trials. Several studies classified ECG finding into major and minor abnormalities by using Minnesota code. Distribution of Minnesota codes in sample subjects. According to a Chinese study [16], major abnormalities have included Minnesota codes as 1-1, 1-2, 4-1, 4-2, 5-1, 5-2, 6-1, 6-2, 7-1, 7-2, 7-4, 8-1 and 8-3 and minor abnormalities were defined as ECGs without any major abnormality and Minnesota codes 1-3, 2-1, 2-2, 3-1, 3-2, 4-3, 5-3, 6-3, 9-1. Belgian interuniversity research on Nutrition and Health (BIRNH) study [17] classified ECG findings into major and minor.

Estimation of diagnostic accuracy

Sensitivity, specificity, positive and negative predictive value, the positive and negative likelihood ratio for SanketLife ECG in comparison to standard 12 lead ECG was calculated for major and ischaemic ECG findings. (Table 3) Overall findings have combinations such as only major code, only minor code and both major and minor codes together. For the analysis purpose, in cases where both major and minor codes were reported together, only major codes were considered. Minor codes and time variant ECG findings such as atrial and ventricular premature beats were not taken into account for matching. Out of total 91 subjects, 53 major codes were accurately detected by SanketLife.
Table 3

Diagnostic accuracy parameters of SanketLife.

ECG typePositiveSanketLife ECG
Total
Negative
12 lead standard ECGPositive53154
Negative03737
Total533891

Sensitivity: 98.15%.

Specificity: 100%.

Positive predictive value: 100%.

Negative predictive value: 97.36%.

Positive likelihood ratio: -

Negative likelihood ratio: 0.02.

Diagnostic accuracy parameters of SanketLife. Sensitivity: 98.15%. Specificity: 100%. Positive predictive value: 100%. Negative predictive value: 97.36%. Positive likelihood ratio: - Negative likelihood ratio: 0.02.

Discussion

Since the time of large room size ECG machines to portable pocket-size ECG devices, electrocardiography has undergone dramatic changes. These transformations helped in minimizing the time period of ‘event to intervention’ drastically in cardiac emergencies. The primary objective of the current study was to assess the diagnostic accuracy of SanketLife ECG with respect to gold standard a 12 lead ECG (GEE2000). Ninety-one patients, with both the ECGs has done, were included in the final analysis. ECG findings were classified according to Minnesota codes by an electrophysiologist. Prevalence of ECG abnormalities (Minnesota codes) was calculated. Among major abnormalities, T wave inversion was the most common finding followed by ST depression and frequent premature beats. An epidemiological study [18], which pooled data from four Belgian populationbased studies, estimated prevalence of ECG abnormalities and found ST depression, T wave inversion as the most prevalent ECG abnormalities.

ST depression and T wave inversion

Minnesota code 4 and 5 represents ST depression and T wave inversion respectively. According to a follow-up study [19] Code 4-1, 4-2, 5-1 and 5-2 were classified as major ECG abnormalities while Code 4-3, 4-4, 5-3 and 5-4 were defined as minor. In our analysis, 10 ECGs reported code 4-1. SanketLife identified similar code in all these subjects. The 4-3 code was reported in 12 subjects by standard 12 lead ECG, among them SanketLife reported it in 10 subjects. Major findings were captured with great accuracy by SanketLife. All 27 subjects with the major category of T wave inversion (5-2), were correctly diagnosed by SanketLife. Among the other 5 subjects two were identified with minor ECG abnormality (5-3) by both types of ECGs. One case reported 5-3 code in SanketLife ECG and not in 12 lead standard ECG, while two subjects reported in standard 122 lead ECG, were not reported by SanketLife. Overall major ECG abnormalities were reported by SanketLife with 100% accuracy.

AV conduction defects and ventricular defects

Among 9 subjects, AV conduction and ventricular defects (Minnesota Code 6 and code 7) were detected by SanketLife as well as by standard 12 lead ECG.

Arrhythmias

A single case of atrial fibrillation (Minnesota code: 8-3-1) was accurately detected by SanketLife. SanketLife reported ventricular premature beats in 7 cases while standard 12 lead ECG in 3 out of 7. There was discordance between two types of ECGs in identifying atrial and ventricular premature beats. A comparative study [20] of 60-s standard ECG vs 24-h Holter monitoring in the detection of ventricular arrhythmias reported that ventricular ectopic beats were present in 17% of standard ECG against 90% of the Holter monitoring among 100 patients admitted for definite acute myocardial infarction. This study concluded standard ECG as an insensitive method for the diagnosis of ventricular arrhythmia. Time is also an important function in the diagnosis of ventricular ectopic. Time-variant nature of these findings is a plausible explanation for the discrepancy between standard ECG and SanketLife in the diagnosis of atrial and ventricular premature beats.

Miscellaneous (Minnesota code 9)

SanketLife ECG correctly diagnosed miscellaneous ECG abnormalities including low QRS amplitude (Minnesota code (MC): 9-1), ST segment elevation (MC:9-2) and hyperacute T waves (MC:9-5). Ten ECGs were classified as normal by both standard and SanketLife ECG.

Conclusion

SanketLife has identified all major ECG abnormalities in high agreement with standard 12 lead ECG. SanketLife has shown high sensitivity in detecting significant abnormalities. Discordance was reported in identifying minor ECG abnormalities and time-dependent atrial and ventricular premature beats where continuous monitoring is a better method. High sensitivity (True positive) and specificity of SanketLife in identifying electrical cardiac abnormalities, its portability coupled with real-time transmission and low cost makes it a suitable candidate for the emergency as well as outpatient based cardiac care in clinical and community settings.

Declaration of competing interest

None.
Table 2

Distribution of Minnesota codes in sample subjects.

Type of ECG abnormalitiesMinnesota CodePrevalence N (%)
Major ECG abnormalities
ST depressionIV1-210 (10.9)
T wave inversionV 1-226 (28.5)
Complete or second-degree AV blockVI 1-23 (3.3)
Complete left or right bundle branch blockVII 1-24 (4.4)
Frequent premature beatsVIII 110 (10.9)
Atrial fibrillationVIII 31 (1.1)
Minor ECG abnormalities
Borderline Q waveI 30 (0)
Left or right axis deviationII 20 (0)
QRS high voltageIII 1-211 (12.1)
Borderline ST segment depressionIV 30 (0.0)
T wave flatteningV 37 (5.5)
QRS low voltageIX 11 (1.1)
Myocardial ischaemia (Ischaemic ECG)
Presence of Q/QS patternsI 1-31 (1.1)
Significant or borderline ST segment depressionIV1-310 (10.9)
Deep or moderate T wave inversionV 1-331 (34.1)
Evidence of complete left bundle branch blockVII 12 (2.2)
  17 in total

1.  Field triage to primary angioplasty combined with emergency department bypass reduces treatment delays and is associated with improved outcome.

Authors:  Steen Carstensen; Greg C I Nelson; Peter S Hansen; Lewis Macken; Stephen Irons; Michael Flynn; Pramesh Kovoor; Soon Y Soo Hoo; Michael R Ward; Helge H Rasmussen
Journal:  Eur Heart J       Date:  2007-08-01       Impact factor: 29.983

Review 2.  Burden of coronary heart disease in India.

Authors:  Rajeev Gupta
Journal:  Indian Heart J       Date:  2005 Nov-Dec

Review 3.  Pre-hospital electrocardiogram triage with telemedicine near halves time to treatment in STEMI: A meta-analysis and meta-regression analysis of non-randomized studies.

Authors:  Natale Daniele Brunetti; Luisa De Gennaro; Michele Correale; Francesco Santoro; Pasquale Caldarola; Antonio Gaglione; Matteo Di Biase
Journal:  Int J Cardiol       Date:  2017-01-08       Impact factor: 4.164

4.  A System of Care for Patients With ST-Segment Elevation Myocardial Infarction in India: The Tamil Nadu-ST-Segment Elevation Myocardial Infarction Program.

Authors:  Thomas Alexander; Ajit S Mullasari; George Joseph; Kumaresan Kannan; Ganesh Veerasekar; Suma M Victor; Colby Ayers; Viji Samuel Thomson; Vijayakumar Subban; Justin Paul Gnanaraj; Jagat Narula; Dharam J Kumbhani; Brahmajee K Nallamothu
Journal:  JAMA Cardiol       Date:  2017-05-01       Impact factor: 14.676

5.  Nationwide improvement of door-to-balloon times in patients with acute ST-segment elevation myocardial infarction requiring primary percutaneous coronary intervention with out-of-hospital 12-lead ECG recording and transmission.

Authors:  Marcus Eng Hock Ong; Aaron Sung Lung Wong; Chong Meng Seet; Swee Guan Teo; Beng Leong Lim; Paul Jau Lueng Ong; Shieh Mei Lai; Sea Hing Ong; Francis Chun Yue Lee; Kim Poh Chan; Venkataraman Anantharaman; Terrance Siang Jin Chua; Pin Pin Pek; Huihua Li
Journal:  Ann Emerg Med       Date:  2012-09-27       Impact factor: 5.721

6.  Standard ECG versus 24-hour Holter monitoring in the detection of ventricular arrhythmias.

Authors:  M Møller
Journal:  Clin Cardiol       Date:  1981 Nov-Dec       Impact factor: 2.882

7.  Prognostic value of ECG findings for total, cardiovascular disease, and coronary heart disease death in men and women.

Authors:  D De Bacquer; G De Backer; M Kornitzer; H Blackburn
Journal:  Heart       Date:  1998-12       Impact factor: 5.994

8.  Effect of potentially modifiable risk factors associated with myocardial infarction in 52 countries (the INTERHEART study): case-control study.

Authors:  Salim Yusuf; Steven Hawken; Stephanie Ounpuu; Tony Dans; Alvaro Avezum; Fernando Lanas; Matthew McQueen; Andrzej Budaj; Prem Pais; John Varigos; Liu Lisheng
Journal:  Lancet       Date:  2004 Sep 11-17       Impact factor: 79.321

9.  The Economic impact of Non-communicable Diseases on households in India.

Authors:  Michael M Engelgau; Anup Karan; Ajay Mahal
Journal:  Global Health       Date:  2012-04-25       Impact factor: 4.185

Review 10.  Futuristic biosensors for cardiac health care: an artificial intelligence approach.

Authors:  Rajat Vashistha; Arun Kumar Dangi; Ashwani Kumar; Deepak Chhabra; Pratyoosh Shukla
Journal:  3 Biotech       Date:  2018-08-03       Impact factor: 2.406

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.