Abstract
Diffusion-weighted echo planar imaging (DW-EPI) suffers from geometric distortion due to low phase-encoding bandwidth. Read-out segmented echo planar imaging (RS-EPI) reduces distortion but residual distortion remains in extreme cases. Additional corrections need to be applied, especially for radiotherapy applications where a high degree of accuracy is needed. In this study the use of magnetic field map corrections are assessed in DW-EPI and RS-EPI, to reduce geometric uncertainty for MRI-guided radiotherapy applications. Magnetic field maps were calculated from gradient echo images and distortion corrections were applied to RS-EPI images. Distortions were assessed in a prostate phantom by comparing to the known geometry, and in vivo using a modified Hausdorff distance metric using a T2-weighted spin echo as ground truth. Across 10 patients, field map-corrected RS-EPI reduced maximum distortion by 5 mm on average compared to DW-EPI (σ = 1.9 mm). Geometric distortions were also reduced significantly using field mapping with RS-EPI, compared to RS-EPI alone (p ≤ 0.05). The increased geometric accuracy of these techniques can potentially allow diffusion-weighted images to be fused with other MR or CT images for radiotherapy treatment purposes.Entities: