Thyroid stimulating hormone values of clinical decisions of hypothyroidism measurement by three different automated immunoassays.

Bosa Mirjanic-Azaric1,2, Ales Jerin3, Zana Radic2.   

Abstract

Background: An accurate measurement of serum thyroid stimulating hormone (TSH) is crucial for a thyroid disorder diagnosis and management. We compared the values of TSH between three automated immunoassays with aims to provide insights into variations in TSH levels that are important for a clinical decision: 2.50 mIU/L, 4.00 mIU/L and 10.00 mIU/L.
Methods: We measured TSH with three different fully automated immunoassays: Abbott (Architect ci8200), Siemens (ADVIA Centaur XP) and Roche (Cobas e411). Serum was collected from 110 patients between August 2018 and January 2019. The results were compared using the Passing-Bablok regression method. Additionally, linear regression coefficients were calculated after logarithmic transformation of data.
Results: Although all three regression coefficients were high (r2 > 0.98), the slopes from Passing-Bablok plots for the correlation of Abbott with either Roche or Siemens were merely 0.66 and 0.73, respectively. The slope for the correlation of Roche and Siemens was 1.11. Consecutively, the results obtained by the Roche and Siemens methods were proportionally higher than those obtained by the Abbott method (38% and 52%, respectively) at all measured ranges.Conclusions: Although immunoassays correlated among themselves, they cannot achieve the same values for clinical decisions for hypothyroidism (clinical requirements). Clinicians should be conscious of these limitations. A harmonisation of the methods is needed to meet clinical requirements and to enable appropriate clinical decisions in cases of hypothyroidism.Likewise, we suggest the introduction of borderline and high risk values of TSH for hypothyroidism depending on immunoassays to avoid misdiagnosis.

Entities:  

Keywords:  Comparison immunoassays of TSH; clinical decisions of subclinical hypothyroidism

Mesh:

Substances:

Year:  2019        PMID: 31841041     DOI: 10.1080/00365513.2019.1703215

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  2 in total

1.  Establishing paediatric reference intervals for thyroid function tests in Croatian population on the Abbott Architect i2000.

Authors:  Adriana Bokulić; Ivana Zec; Domagoj Marijančević; Sanja Goreta; Marija Požgaj Šepec; Lavinia La Grasta Sabolić
Journal:  Biochem Med (Zagreb)       Date:  2021-08-05       Impact factor: 2.313

2.  Indirect estimation of reference intervals for thyroid parameters using advia centaur XP analyzer.

Authors:  Bosa Mirjanić-Azarić; Neda Milinković; Natasa Bogavac-Stanojević; Sanja Avram; Tanja Stojaković-Jelisavac; Darja Stojanović
Journal:  J Med Biochem       Date:  2022-04-08       Impact factor: 2.157

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.