In Vivo Evaluation of 3 Electronic Apex Locators: Root ZX Mini, Apex ID, and Propex Pixi.

Gabriela Serna-Peña1, Susana Gomes-Azevedo2, Jorge Flores-Treviño1, Elizabeth Madla-Cruz1, Idalia Rodríguez-Delgado1, Gustavo Martínez-González3.   

Abstract

INTRODUCTION: The aim of this in vivo study was to compare the accuracy of 3 electronic apex locators (EALs) (Root ZX mini [J Morita Corp, Tokyo, Japan], Apex ID [SybronEndo, Glendora, CA], and Propex Pixi [Dentsply Maillefer, Ballaigues, Switzerland]) to determine the working length.
METHODS: Thirty single-rooted human teeth that were scheduled for extraction were selected for the study. Electronic measurements were performed with the 3 EALs. After the teeth had been extracted, a #10 K-file was used to determine the actual working length, which was established at 0.5 mm short of the major foramen. The data were statistically analyzed with analysis of variance (α = 0.05).
RESULTS: No significant differences were found among the experimental groups (P > .05). The mean distance from the actual working length to the file tip was 0.163 ± 0.032 mm when Root ZX mini was used, 0.343 ± 0.032 mm for Propex Pixi , and 0.012 ± 0.008 mm for Apex ID.
CONCLUSIONS: Under the in vivo conditions of this study, no statistically significant differences were observed among the 3 EALs.
Copyright © 2019 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Apex ID; Propex Pixi; Root ZX mini; electronic apex locator; working length

Mesh:

Year:  2019        PMID: 31839411     DOI: 10.1016/j.joen.2019.10.035

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  3 in total

1.  Effect of Orthodontic Brackets on the Accuracy of Apex Locators: A Pilot Study.

Authors:  Özgür Genç Şen; Özgür İlke Ulusoy; Yelda Nayır Paltun; Çağrı Ulusoy
Journal:  Biomed Res Int       Date:  2021-04-24       Impact factor: 3.411

Review 2.  Artificial Intelligence in Dentistry: Past, Present, and Future.

Authors:  Paridhi Agrawal; Pradnya Nikhade
Journal:  Cureus       Date:  2022-07-28

Review 3.  Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.

Authors:  Kaveh Nasiri; Karl-Thomas Wrbas
Journal:  Saudi Dent J       Date:  2021-09-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.