Τhe AFX unibody bifurcated unibody aortic endograft for the treatment of abdominal aortic aneurysms: current evidence and future perspectives.

Efstratios Georgakarakos1, Georgios Ioannidis2, Andreas Koutsoumpelis1, Nikolaos Papatheodorou3, Christos Argyriou1, Konstantinos Spanos4, Athanasios D Giannoukas4, George S Georgiadis1.   

Abstract

Introduction: AFX is a bifurcated unibody aortic endograft for the treatment of Abdominal Aortic aneurysms. It consists of an inner metallic endoskeleton with multiple metallic struts covered by a polytetrafluoroethylene graft fabric. The endoskeleton is sutured to the outer fabric only at the proximal and distal ends. The unique design of AFX aims at fixation onto the aortic bifurcation while a proximal cuff ensures sealing at the infrarenal level.Areas covered: Since this endograft design by Endologix has undergone significant changes over the last years, the aim of this article is to present its unique structure and deployment method and discuss the relevant clinical results as well as reported complications and associated concerns.Expert commentary: the AFX stent-graft exhibits very satisfactory clinical mid-term results in abdominal aortic aneurysms treated within the instructions-for-use. Its 'active-seal' concept of infrarenal fixation stemming from the loose conjugation of the fabric material to the endoskeleton can accommodate efficiently to challenging necks with thrombus or morphological irregularities, thereby extending the anatomical sealing zone without exerting significant radial outward force. Long-term results are needed to validate the promising performance of AFX.

Entities:  

Keywords:  AFX; Endologix; abdominal aortic aneurysm; aorta; endovascular; stent-graft; unibody bifurcated endograft

Mesh:

Year:  2019        PMID: 31833437     DOI: 10.1080/17434440.2020.1704254

Source DB:  PubMed          Journal:  Expert Rev Med Devices        ISSN: 1743-4440            Impact factor:   3.166


  2 in total

1.  Use of the AFX Stent Graft in Patients with Extremely Narrow Aortic Bifurcation: A Multicenter Retrospective Study.

Authors:  M U Wagenhäuser; N Floros; E Nikitina; J Mulorz; K M Balzer; S Goulas; M Petrich; P Dueppers; F Simon; H Schelzig; A Oberhuber
Journal:  Int J Vasc Med       Date:  2021-10-04

2.  Surgery for late type Ia/IIIb endoleak from a fabric tear and stent fracture of AFX2 stent graft.

Authors:  Simone Cuozzo; Francesca Miceli; Antonio Marzano; Ombretta Martinelli; Roberto Gattuso; Enrico Sbarigia
Journal:  J Vasc Surg Cases Innov Tech       Date:  2022-07-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.