Literature DB >> 31828243

Use of a Combined Laryngo-Bronchoscopy Approach in Difficult Airways Management: A Pilot Simulation Study.

Filippo Sanfilippo1, Francesco Sgalambro2, Giuseppe Chiaramonte3, Cristina Santonocito1, Gaetano Burgio1, Antonio Arcadipane1.   

Abstract

OBJECTIVE: There are several airway devices available for difficult tracheal intubation (DTI) management, but the failure rate remains high. The use of laryngoscopy to facilitate the fibreoptic-bronchoscope intubation (CLBI) has been increasingly reported in DTI situations, but it has not been formally studied yet.
METHODS: We designed a single-centre simulation study on DTI (neck rigidity and tongue oedema) comparing three techniques: direct laryngoscopy (DL), video-laryngoscopy (VLS) and CLBI. Eighteen anaesthesiologists naïve to VLS/CLBI approaches, participated in the study. The primary outcome was the intubation rate at the first attempt. Secondary outcomes were an overall time-to-intubate (TTI) and time-to-ventilate (TTV), success at the second and third attempt and ease of intubation as evaluated by a subjective 5-point Likert scale.
RESULTS: The CLBI technique had a higher success rate at the first attempt than DL (66% vs 22%, p=0.007), while VLS did not (44%, p=0.16). A trend towards higher success at the third attempt was found for both VLS and CLBI vs DL (p=0.07 and p=0.06, respectively). The VLS had a shorter overall TTV than DL (88±60 vs 121±59 sec, respectively, p=0.04) and a trend towards a shorter TTI (81±61 vs 116±64 sec, respectively, p=0.06). The CLBI approach showed a non-significantly lower TTI/TTV as compared to DL (p=0.10 and p=0.16, respectively). Anaesthesiologists judged that the intubation with VLS (3.7±1.0) and CLBI (3.8±1.0) was easier than with DL (1.7±0.8, both p<0.001).
CONCLUSION: In a simulated DTI scenario, CLBI had a higher success rate at the first attempt than DL, while VLS did not. By the third attempt, both rescue techniques had a trend towards a higher success rate than DL. The CLBI technique seems a promising alternative for the management of DTI. © Copyright 2019 by Turkish Anaesthesiology and Intensive Care Society.

Entities:  

Keywords:  Difficult tracheal intubation; fibreoptic intubation; manikin; tongue oedema; video-laryngoscope

Year:  2019        PMID: 31828243      PMCID: PMC6886816          DOI: 10.5152/TJAR.2019.99234

Source DB:  PubMed          Journal:  Turk J Anaesthesiol Reanim        ISSN: 2149-276X


  29 in total

1.  Bougie-related airway trauma: dangers of the hold-up sign.

Authors:  B A Marson; E Anderson; A R Wilkes; I Hodzovic
Journal:  Anaesthesia       Date:  2014-03       Impact factor: 6.955

2.  A combined rigid videolaryngoscopy-flexible fibrescopy intubation technique under general anesthesia.

Authors:  Nicolas Greib; Nathalie Stojeba; W Allister Dow; John Henderson; Pierre A Diemunsch
Journal:  Can J Anaesth       Date:  2007-06       Impact factor: 5.063

3.  Complications of awake fibreoptic intubation without sedation in 200 healthy anaesthetists attending a training course.

Authors:  N M Woodall; R J Harwood; G L Barker
Journal:  Br J Anaesth       Date:  2008-04-19       Impact factor: 9.166

4.  Learning endotracheal intubation using a novel videolaryngoscope improves intubation skills of medical students.

Authors:  Frank Herbstreit; Philipp Fassbender; Helge Haberl; Clemens Kehren; Jürgen Peters
Journal:  Anesth Analg       Date:  2011-06-16       Impact factor: 5.108

5.  Combined use of an Airtraq® optical laryngoscope, Airtraq video camera, Airtraq wireless monitor, and a fibreoptic bronchoscope after failed tracheal intubation.

Authors:  Manuel Angel Gómez-Ríos; Laura Nieto Serradilla
Journal:  Can J Anaesth       Date:  2011-01-22       Impact factor: 5.063

6.  Complaints related to respiratory events in anaesthesia and intensive care medicine from 1994 to 1998 in Denmark.

Authors:  C Rosenstock; J Møller; A Hauberg
Journal:  Acta Anaesthesiol Scand       Date:  2001-01       Impact factor: 2.105

7.  First experience of the use of the C-MAC PM videolaryngoscope in a clinical setting by anesthetic nurses: a comparison with anesthetists.

Authors:  Z Kontouli; K Stroumpoulis; E Halari-baradaki; L Papadimitriou; N Iacovidou; T Xanthos
Journal:  Acta Anaesthesiol Belg       Date:  2013

8.  Comparison between bougies and stylets for simulated tracheal intubation with the C-MAC D-blade videolaryngoscope.

Authors:  Bisanth Batuwitage; Andrew McDonald; Koji Nishikawa; Daniel Lythgoe; Simon Mercer; Peter Charters
Journal:  Eur J Anaesthesiol       Date:  2015-06       Impact factor: 4.330

9.  Management of the difficult airway: a closed claims analysis.

Authors:  Gene N Peterson; Karen B Domino; Robert A Caplan; Karen L Posner; Lorri A Lee; Frederick W Cheney
Journal:  Anesthesiology       Date:  2005-07       Impact factor: 7.892

10.  Difficult Airway Society 2015 guidelines for management of unanticipated difficult intubation in adults.

Authors:  C Frerk; V S Mitchell; A F McNarry; C Mendonca; R Bhagrath; A Patel; E P O'Sullivan; N M Woodall; I Ahmad
Journal:  Br J Anaesth       Date:  2015-11-10       Impact factor: 9.166

View more
  1 in total

1.  Tracheobronchial intubation using flexible bronchoscopy in children with Pierre Robin sequence: Nursing considerations for complications.

Authors:  Ying-Long Ye; Cai-Feng Zhang; Li-Zhen Xu; Hui-Feng Fan; Jun-Zheng Peng; Gen Lu; Xiao-Yin Hu
Journal:  World J Clin Cases       Date:  2022-07-06       Impact factor: 1.534

  1 in total